Titanium Trochanteric Fixation Nail (TFN)System recallZ-0452-2014 · Synthes USA HQ, Inc. · initiated Jul 30, 2013
Class II Terminated All recalls posted in December 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Synthes USA HQ, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Two lots of Titanium Trochanteric Fixation Nail (TFN) System are missing the second bend in the nail.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Synthes sent an Urgent Notice: Medical Device Recall letter dated July 30, 2013. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to call Synthes at 1-800-479-6329 if they have any of the affeccted product to obtain a Return Authorization Number, Complete the Verification Section of the letter by checking the appropriate box indicating the affected product has been located. Return the Verification Form with the product to: Credit/Returns Synthes 1101 Synthes Avenue Monument, CO 80132. Customers should complete and return the Verification Frorm even if they do not have any of the affected product and return by fax to 1-888-386-2077 or Scan/email: [email protected]. Customers with questions were instructed to call 610-719-5450 or contact their Synthes Trauma Sales Consultant. For questions regarding this recall call 610-719-5000.
Recall details
- Recall number
- Z-0452-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 30, 2013
- Posted by the FDA
- Dec 5, 2013
- Recalling firm
- Synthes USA HQ, Inc.
- Quantity
- 17
- Product code
- HSB
- FDA submissions
- K011857
- Distribution
- Nationwide Distribution including IL, CA, AZ, FL, TX, UT, VA, NY, NM, and MT.
Product description
Titanium Trochanteric Fixation Nail (TFN)System Intended to treat stable and unstable fractures of the proximal femur.
Code information (models, lots, serial numbers)
Part # 456.391S with lot numbers 5536930, exp. 5/20/2016, and 7161401, exp. 12/31/2021.
Other devices under the same FDA submission
44 devices cleared under K011857 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA 456.305Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES9 reports
- NA 456.303Synthes GmbH · Screw, fixation, bone7 reports
- NA 456.315SSynthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES7 reports
- NA 456.318SSynthes GmbH · Screw, fixation, bone6 reports
- NA 456.306Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES5 reports
- NA 456.302Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES4 reports
- NA 456.304SSynthes GmbH · Screw, fixation, bone4 reports
- NA 456.306SSynthes GmbH · Screw, fixation, bone4 reports
- NA 456.322SSynthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES4 reports
- NA 456.420SSynthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES4 reports
- NA 456.305SSynthes GmbH · Screw, fixation, bone3 reports
- NA 456.415SSynthes GmbH · Screw, fixation, bone3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page