Synthes Titanium Cannulated Humeral Nail recallZ-0407-2014 · Synthes USA HQ, Inc. · initiated Jul 30, 2013
Class II Terminated All recalls posted in November 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Synthes USA HQ, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The packaging incorrectly depicts the Synthes Titanium Cannulated Humeral Nail as a 3 distal hole nail instead of a 2 distal locking hole nail.
FDA root cause category: Pending
What the company is doing
Synthes sent an Urgent Notice: Medical Device Recall letter dated July 30, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to call Synthes at 1-800-479-6329 to obtain a Return Authorization Number, complete the Verification Section and return to Credit/Returns, Synthes, 1101 Synthes Avenue, Monument, CO 80132. If customers do not have any affected product they should still complete Verification Form and return to Synthes by fax to 1-888-731-7954 or Scan/email: [email protected]. Customers with questions were instructed to call 610-719-5450. For questions regarding this recall call 610-519-5000.
Recall details
- Recall number
- Z-0407-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 30, 2013
- Posted by the FDA
- Nov 27, 2013
- Recalling firm
- Synthes USA HQ, Inc.
- Quantity
- 14
- Product code
- JDS
- FDA submissions
- K033071
- Distribution
- Worldwide Distribution - USA (nationwide) including SC, IA, NC, TX, KY, VA, AL, PA, and Internationally to Canada.
Product description
Synthes Titanium Cannulated Humeral Nail Intended to aid in the alignment and stabilization of humeral fractures
Code information (models, lots, serial numbers)
Part # Lot # 04.001.620S 7115414 04.001.620S 7115417 04.001.620S 7115509 04.001.620S 7115510 04.001.622S 7115533 04.001.636S 7117902 04.001.636S 7117916 04.001.638S 7117924 04.001.640S 7119089 04.001.640S 7119107 04.001.642S 7117995 04.001.642S 7119125 04.001.644S 7117998 04.001.218S 7327816 04.001.444S 7379241 04.001.620s 7265517
Other devices under the same FDA submission
12 devices cleared under K033071 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- EXPert Nail 04.001.002Synthes GmbH · NAIL, FIXATION, BONE2 reports
- EXPert Nail 04.001.428SSynthes GmbH · NAIL, FIXATION, BONE2 reports
- EXPert Nail 04.001.000Synthes GmbH · NAIL, FIXATION, BONE1 report
- NA 04.001.007SSynthes GmbH · NAIL, FIXATION, BONE1 report
- NA 04.001.008SSynthes GmbH · NAIL, FIXATION, BONE1 report
- NA 04.001.009SSynthes GmbH · NAIL, FIXATION, BONE1 report
- NA 04.001.010SSynthes GmbH · NAIL, FIXATION, BONE1 report
- EXPert Nail 04.001.224SSynthes GmbH · NAIL, FIXATION, BONE1 report
- EXPert Nail 04.001.234SSynthes GmbH · NAIL, FIXATION, BONE1 report
- EXPert Nail 04.001.236SSynthes GmbH · NAIL, FIXATION, BONE1 report
- EXPert Nail 04.001.410SSynthes GmbH · NAIL, FIXATION, BONE1 report
- EXPert Nail 04.001.440SSynthes GmbH · NAIL, FIXATION, BONE1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page