Large Buttress Compression Nut For Blade Guide Sleeve TFN recallZ-0387-2014 · Synthes USA HQ, Inc. · initiated Jul 31, 2013
Class II Terminated All recalls posted in November 2013
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Synthes USA HQ, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Large Buttress Compression Nut For Blade Guide Sleeve TFN was recalled because of complaints that the compression nut did not thread completely onto the blade guide sleeve and became stuck.
FDA root cause category: Process control
What the company is doing
Synthes sent an Urgent Notice: Medical Device Recall letter dated July 31, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Action to be taken: Our records indicate you received this product. Synthesis requesting that the product be returned: If you DO have any of the identified devices, please take the following steps: "Call Synthes at 1-800-479-6329 to obtain a Return Authorization Number. "Complete the Verification Section at the end of this letter by checking the appropriate box indicating affected product has been located. Also, please indicate the number of devices found and note the Return Authorization Number. Please include your name, title, telephone number and signature in the spaces provided. "Return the Verification Form {page 3 of this letter) with the product to: "Credit/Returns, Synthes, .1101 Synthes Avenue, Monument, CO 80132. If you DO NOT have the identified product, please take the following steps: "Complete the attached Verification Section at the end of this letter by checking the appropriate box indication that no affected product has been located. Please include your name, title, telephone number and signature in the spaces provided this return documentation acknowledges your receipt of medical device removal information. "Return the documents to Synthes by. " Fax: 1-888-731-7952 or " Scan/email: [email protected] If you have any questions, please call 610-719-5450 or contact your Synthes Sales Consultant. Thank you for your attention to this issue
Recall details
- Recall number
- Z-0387-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 31, 2013
- Posted by the FDA
- Nov 25, 2013
- Recalling firm
- Synthes USA HQ, Inc.
- Quantity
- 135
- Product code
- HSB
- FDA submissions
- K011857
- Distribution
- US Distribution including the states of TX, PA, NY, GA, CO, IN, AI, AR, IL, MI, TN, CA, AZ, MS and VA.
Product description
Large Buttress Compression Nut For Blade Guide Sleeve TFN The Synthes Large Buttress Compression Nut For Blade Guide Sleeve TFN is intended to treat stable and unstable pertrochanteric fractures, intertrochanteric fractures, basal neck fractures, and combinations there of.
Code information (models, lots, serial numbers)
Part # Lot # SD357.371 5734551 6083065 6538811 6631975 6703191 6703190 6841538
Other devices under the same FDA submission
44 devices cleared under K011857 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA 456.305Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES9 reports
- NA 456.303Synthes GmbH · Screw, fixation, bone7 reports
- NA 456.315SSynthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES7 reports
- NA 456.318SSynthes GmbH · Screw, fixation, bone6 reports
- NA 456.306Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES5 reports
- NA 456.302Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES4 reports
- NA 456.304SSynthes GmbH · Screw, fixation, bone4 reports
- NA 456.306SSynthes GmbH · Screw, fixation, bone4 reports
- NA 456.322SSynthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES4 reports
- NA 456.420SSynthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES4 reports
- NA 456.305SSynthes GmbH · Screw, fixation, bone3 reports
- NA 456.415SSynthes GmbH · Screw, fixation, bone3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page