Fresenius 2008 Series Hemodialysis Machines recallZ-0629-2014 · Fresenius Medical Care Holdings, Inc. · initiated Nov 12, 2013
Class II Terminated All recalls posted in January 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Fresenius Medical Care Holdings, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
2008 Series Hemodialysis Machines complaints of saline bags inappropriately filling during recirculation and setup
FDA root cause category: Under Investigation by firm
What the company is doing
Fresenius Medical issued a Product Notification letter dated November 12, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers who encounter a saline bag that is inappropriately filling, should discontinue the setup and remove the machine from service until the drain line and drain-associated hydraulic components can be checed by a qualified service technician. Customers were also instructed to ensure that the hemodialysis machine is installed and operated in compliance with the Machine Specification section of their operator's manual, including the specification located in the Water/Drain Section. Customers with questions were instructed to call 1-800-227-2572. For questions regarding this recall call 800-662-1237.
Recall details
- Recall number
- Z-0629-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 12, 2013
- Posted by the FDA
- Jan 6, 2014
- Recalling firm
- Fresenius Medical Care Holdings, Inc.
- Quantity
- 111,504 units US and 3,638 Canada and Mexico-
- Product code
- KDI
- FDA submissions
- K070049, K093902, K921456, K994267
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally Mexico and Canada.
Product description
Fresenius 2008 Series Hemodialysis Machines: Models: 2008H, 2008K, 2008K2, 2008k@Home, 2008T The 2008H Hemodialysis machine is indicated for acute and chronic dialysis therapy. Hemodialysis is indicated for patients with acute or chronic renal failure, when conservative therapy is judged to be inadequate. The dialysate delivery machine is used for hemodialysis in hospital, dialysis centers, and at home. The 2008K, 20081(2 and 2008T hemodialysis machines are designed to perform hemodialysis in hospitals and dialysis clinics. They can be used for patients suffering chronic or acute renal failure. The 2008K, 2008K2 and 2008T are designed to provide hemodialysis treatment by controlling and monitoring both the dialysate and extracorporeal blood circuits. The Fresenius 2008K~home is indicated for acute and chronic dialysis therapy in an acute or chronic facility. The 2008K~home is also indicated for hemodialysis in the home and must be observed by a trained and qualified person as prescribed by their physician.
Code information (models, lots, serial numbers)
All serial numbers
Other devices under the same FDA submission
2 devices cleared under K070049, K093902, K921456, K994267 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- 2008® Series Hemodialysis Machines 190828Fresenius Usa, Inc. · Hemodialysis system for home use2 recalls8 reports
- CombiSet 03-2962-3Fresenius Medical Care Holdings, Inc. · Hemodialysis system for home use3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.