PedFuse Reset Screw recallZ-0544-2014 · SpineFrontier, Inc. · initiated Oct 18, 2013

Class II Terminated All recalls posted in December 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from SpineFrontier, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During routine servicing, it was discovered that one 8mmx55mm Reset screw was assembled with an 8mm tulip head and a 9mm screw. The implant should have consisted of an 8mm tulip head and an 8mm screw. A recall was initiated to remove all remaining screws of the same lot from the field to investigate the nonconformity.

FDA root cause category: Software design (manufacturing process)

What the company is doing

SpineFrontier sent an Voluntary Recall Notification dated November 5,2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. All affected implants are to be returned to SpineFrontier for inspection. SpineFrontier is requiring confirmation that you have received this notice and that you will initiate the immediate return of the implants listed above. IMMEDIATE RESPONSE REQUIRED. Please feel free to contact us at 978-232-3990 Monday through Friday 8:30AM to 5:30PM EST with any questions

Recall details

Recall number
Z-0544-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 18, 2013
Posted by the FDA
Dec 19, 2013
Recalling firm
SpineFrontier, Inc.
Quantity
35 Devices
Product code
MNI
FDA submissions
K123164
Distribution
US Distribution to VA and TX.

Product description

PedFuse Reset Screw; Model: 01-80115-55 The PedFuse Reset pedicle screw implants intended to be used to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of the thoracic, lumbar, and/or sacral spine.

Code information (models, lots, serial numbers)

Lot/Serial Numbers: 357305-000

Other devices under the same FDA submission

1 device cleared under K123164 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.