Abbott i-STAT PT/INR cartridges Abbott Point of Care Inc.… recallZ-0451-2014 · Abbott Point Of Care Inc. · initiated Oct 28, 2013

Class II Terminated All recalls posted in December 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Point Of Care Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Abbott Point of Care Inc (APOC) has determined that i-Stat PT/INR cartridges have the potential to exhibit incorrectly elevated INR.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Abbott Point of Care Inc. (APOC) issued an Urgent Recall Notice on October 28, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of cartridges from PT/INR lots in this letter. All unused cartridges would be credited and are to be returned as instructed on the enclosed Business Reply Card. If customers provided cartridges from the lot number listed above to another institution they were instructed to provide a copy of this letter to them. Customers with questions were instructerd to contact Abbott Point of Care Technical at 1-800-366-8020, Option 1. For questions regarding this recall call 800-366-8020.

Recall details

Recall number
Z-0451-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 28, 2013
Posted by the FDA
Dec 4, 2013
Recalling firm
Abbott Point Of Care Inc.
Quantity
794,928 units US; 119,472 units OUS
Product code
GJS
FDA submissions
K020355
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia, Brazil, Chile, China, Finland, France, Germany, Greece, Hong Kong, Ireland, Italy, South Korea, Kuwait, Mexico, Netherland, New Zealand, Nigeria, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, south Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, UAE, and United Kingdom,

Product description

Abbott i-STAT PT/INR cartridges Abbott Point of Care Inc., Princeton, NJ 08540 The i-Stat PT a prothrombin time test is useful for monitoring patients receiving oral anticoagulation therapy such as Coumadin or Warfarin. The cartridge is to be used with the i-Stat 1 Analyzer and as part of the i-Stat System, the PT/INR test is to be used by trained health care professionals in accordance with a facility's policies and procedures.

Code information (models, lots, serial numbers)

K020355  List number 04J50-01, 04J50-02; 03P89-24  Lot number R13149A; R13151; R13151A and cartridge lots from C13139 up to and including C13270A

Other devices under the same FDA submission

1 device cleared under K020355 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.