Atrium Oasis Dry Suction Water Seal ATS (Autotransfusion)… recallZ-0740-2014 · Atrium Medical Corporation · initiated Nov 19, 2013
Class II Terminated All recalls posted in January 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Atrium Medical Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Chest Drain tubing of of the ATS Blood recovery may leak or disconnect
FDA root cause category: Device Design
What the company is doing
Atrium Medical issued notification letters to direct consignees by UPS next day on 19 November 2013, and indirect consignees (distributors customers) by UPS next day or USPS priority (for customers with a PO Box) on 27 November 2013. The Letter explain the issue and the steps that should be taken. The customer will be offered replacement with unaffected product or credit. The firm describes Procedures for Emergency Use. The firm identifies potential product substitutions. These products are the same drainage product without the auto-transfusion line. International Distributors. International distributors provided with the Recall Authorization' package and the customer letter. Accounts are requested to complete the Recall Reply Form acknowledging receipt of the notification. Contact Atrium Medical Customer Service at 1-800-370-7899, Monday through Friday between 9:00 am to 5:00 pm.
Recall details
- Recall number
- Z-0740-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 19, 2013
- Posted by the FDA
- Jan 14, 2014
- Recalling firm
- Atrium Medical Corporation
- Quantity
- 42,699 units
- Product code
- CAC
- FDA submissions
- K043140, K043582
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Australia, Canada, Austria, Ireland, Romania, South Africa, Colombia, Lebanon Netherlands, Jordan, Russia, and Taiwan.
Product description
Atrium Oasis Dry Suction Water Seal ATS (Autotransfusion) Blood Recovery System Chest Drain with in-line connectors Product Code: 3650-100. To evacuate air and/or fluid from the chest cavity or mediastinum.
Code information (models, lots, serial numbers)
Lot number starting with 10883097 and the ending lot number 10913346. (All units in between and including these two numbers)
Other devices under the same FDA submission
12 devices cleared under K043140, K043582 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- OCEAN 2002-000Atrium Medical Corporation · BOTTLE, COLLECTION, VACUUM3 recalls44 reports
- OCEAN 2002-400Atrium Medical Corporation · BOTTLE, COLLECTION, VACUUM2 recalls18 reports
- OCEAN 2002-300Atrium Medical Corporation · BOTTLE, COLLECTION, VACUUM1 recall10 reports
- OCEAN 2012-320Atrium Medical Corporation · BOTTLE, COLLECTION, VACUUM1 recall3 reports
- OASIS 3620-100Atrium Medical Corporation · Bottle, Collection, Vacuum51 reports
- OCEAN 2002-100Atrium Medical Corporation · BOTTLE, COLLECTION, VACUUM1 recall0 reports
- OCEAN 2002-040Atrium Medical Corporation · BOTTLE, COLLECTION, VACUUM1 recall0 reports
- OCEAN 2020-000Atrium Medical Corporation · BOTTLE, COLLECTION, VACUUM32 reports
- OASIS 3612-400Atrium Medical Corporation · Bottle, Collection, Vacuum3 reports
- OCEAN 2020-100Atrium Medical Corporation · BOTTLE, COLLECTION, VACUUM1 report
- OCEAN 2050-300Atrium Medical Corporation · Apparatus, Autotransfusion1 report
- OASIS 3600-150Atrium Medical Corporation · APPARATUS, AUTOTRANSFUSION1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 66788).
- Z-0735-2014 · TerminatedAtrium Ocean Water Seal Chest Drain ATS (Autotransfusion) Blood Recovery System with stopcock and in-line connectors; Product Code: 2050-000. To evacuate air and/or fluid from th…
- Z-0736-2014 · TerminatedAtrium Ocean Water Seal Chest Drain ATS Blood Recovery System with stopcock and in-line connectors Sterile fluid path packaging Product Code: 2050-070. To evacuate air and/or f…
- Z-0737-2014 · TerminatedAtrium Ocean Water Seal Chest Drain ATS (Autotransfusion) Blood Recovery System with stopcock Blood Recovery System; Product Code: 2050-300. To evacuate air and/or fluid from the …
- Z-0738-2014 · TerminatedAtrium Ocean Water Seal Chest Drain ATS (Autotransfusion) Blood Recovery System Dual patient lines with stopcock and in-line connectors Product Code: 2052-000. To evacuate air a…
- Z-0739-2014 · TerminatedAtrium Ocean Water Seal Chest Drain ATS (Autotransfusion) Blood Recovery System Dual patient lines with stopcock; .Product Code: 2052-300. To evacuate air and/or fluid from the …
- Z-0741-2014 · TerminatedAtrium Oasis Dry Suction Water Seal ATS (Autotransfusion) Blood Recovery System Chest Drain Dual patient lines; Product Code: 3652-100. To evacuate air and/or fluid from the ch…
- Z-0742-2014 · TerminatedAtrium Express Dry Seal ATS (Autotransfusion) Blood Recovery System Chest Drain; Product Code: 4050-100N. To evacuate air and/or fluid from the chest cavity or mediastinum.
- Z-0743-2014 · TerminatedAtrium Express Dry Seal ATS (Autotransfusion) Blood Recovery System Chest Drain with Prefilled Air Leak Monitor; Product Code: 4050-100P. To evacuate air and/or fluid from the …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.