STERRAD NX, Product Code: 10033 recallZ-0403-2014 · Advanced Sterilization Products · initiated Nov 11, 2013

Class II Terminated All recalls posted in November 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Advanced Sterilization Products or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Advanced Sterilization Products (ASP) is recalling the STERRAD NX and STERRAD 100NX sterilization systems because ASP has determined that the bronchoscopes may contain a polyurethane material in the suction channel location that is not listed in the current claims for interior channels (lumens) for processing in STERRAD systems.

FDA root cause category: Labeling design

What the company is doing

Advanced Sterilization Products sent an Urgent Medical Device Field Safety Notificaiton letter on November 11, 2013, to all affected customers. The letter informed customers of the problems identified and the actions to be taken. Customers were instructed to provide the notice to anyone in their facility that needs to be informed and post this notice where appropriate. Customers were instructed to contact ASP at 1-888-783-7723 for any suspected problems. ASP is collaborating with KARL STORZ Endoskope to determine the sterilization impact of the suction port lumen material. Customers are instructed to contact KARL STORZ Endoskope Technical Support at (800)421-0837, ext 5350 for alternative reprocessing instructions. For questions regarding this recall call 949-453-6400.

Recall details

Recall number
Z-0403-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Nov 11, 2013
Posted by the FDA
Nov 26, 2013
Recalling firm
Advanced Sterilization Products
Quantity
Total 4,381 units
Product code
MLR
FDA submissions
K042116
Distribution
Nationwide Distribution

Product description

STERRAD NX, Product Code: 10033 The STERRAD Steri lization System is a low temperature general purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.

Code information (models, lots, serial numbers)

Please refer to consignee list

Other devices under the same FDA submission

1 device cleared under K042116 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 66815).