Why it was recalled

Horiba Medical is recalling the ABX PENTRA N Control because it has been updated to include a revised Target Value assignment for Potassium.

FDA root cause category: Under Investigation by firm

What the company is doing

Horiba sent an customer notification letter dated November 32, 2013, to all affected customers. The letter was sent via email to all customers who purchased the ABX PENTRA 400. HORIBA Medical is informing all ABX PENTRA N Control customers running the Potassium-E assay on the ABX PENTRA 400 with I.S.E. Module that the Target Value and Target Range for Potassium has been updated. The letter informs the customers of the actions to be taken. Customers are instructed to complete and return the enclosed response form within 10 days by fax (949) 453-0600 Customers with questions are instructed to contact their local HORIBA Medical representative. For further questions please call (949) 453-0500 X 208

Recall details

Recall number
Z-0904-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Nov 21, 2013
Posted by the FDA
Jan 31, 2014
Recalling firm
Horiba Instruments, Inc dba Horiba Medical
Quantity
698 units total (80 units in US)
Product code
JJY
FDA submissions
K052007
Distribution
Worldwide Distribution - USA (nationwide)

Product description

ABX PENTRA N Control, Model No. A11A01653 The ABX PENTRA N Control is for use in quality control by monitoring accuracy and precision for the quantitative methods as specified in the enclosed annex.

Code information (models, lots, serial numbers)

Lot No. 1300801

Other devices under the same FDA submission

1 device cleared under K052007 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.