Siemens ADVIA 1200 and ADVIA 1650 Chemistry Systems recallZ-0800-2014 · Siemens Healthcare Diagnostics, Inc. · initiated Dec 17, 2013

Class II Terminated All recalls posted in January 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An overflow flag //// is displayed in the result column for a test when the algorithm for the test cannot calculate a result or when the concentration cannot be determined because it is outside the absorbance limits, and it is possible for the user to misinterpret or edit the result.

FDA root cause category: Software design

What the company is doing

The distribution of the Urgent Medical Device Correction letter scheduled to begin on December 17, 2013. x Field service personnel will be notified of the Urgent Medical Device Correction Letter via the Siemens Knowledge Base, as well as information regarding placing labels on the customer system to remind the customer that ///// is not a reportable result. x The Customers were instructed to contact Siemens Healthcare Diagnostics Technical Solution Center if they have technical questions regarding this issue. x Customers are requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Correction.

Recall details

Recall number
Z-0800-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Dec 17, 2013
Posted by the FDA
Jan 21, 2014
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
1563
Product code
JJE
FDA submissions
K990346
Distribution
Worldwide Distribution. USA nationwide and the country of Canada.

Product description

Siemens ADVIA 1200 and ADVIA 1650 Chemistry Systems. Automated, clinical chemistry analyzers that can run tests on human serum, plasma, or urine.

Code information (models, lots, serial numbers)

ADVIA 1200 with software versions: V2.00 and V2.01  ADVIA 1650 with software versions V3.52 and V4.01

Other devices under the same FDA submission

2 devices cleared under K990346 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.