VERIFY Dual Species Self-Contained Biological Indicator… recallZ-0605-2014 · Steris Corporation · initiated Sep 10, 2013

Class II Terminated All recalls posted in December 2013

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Steris Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

STERIS has identified that the population of the G. stearothermophilus spores that monitor STEAM sterilization cycles does not meet specifications for Lot # 131004 of the Verify Dual Species Self-Contained Biological Indicators. As a result the spore count on the affected product is lower than the labeled minimum and the product may incorrectly indicate that a 5-log spore kill was achieved.

FDA root cause category: Use error

What the company is doing

On September 10, 2013, the firm sent Urgent Voluntary Recall Notices to their customers.

Recall details

Recall number
Z-0605-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Sep 10, 2013
Posted by the FDA
Dec 28, 2013
Recalling firm
Steris Corporation
Quantity
1,238 boxes
Product code
FRC
FDA submissions
K083665, K854631, K854902
Distribution
Worldwide Distribution-USA (nationwide) including the states of AK, Al, AR, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and WY, and the countries of BAHRAIN, CANADA, COLUMBIA, GERMANY, ITALY, MALAYSIA, PHILIPPINES, ROMANIA, SPAIN, and THE NETHERLANDS.

Product description

VERIFY Dual Species Self-Contained Biological Indicator, Distributed by STERIS Corporation, Mentor, OH. Intended for use in installation testing and routine monitoring of steam sterilization and ethylene oxide sterilization processes.

Code information (models, lots, serial numbers)

Lot #131004 Model #'s: S3060, S3061, S3065, S3069, LCB006 & LCB007.

Other devices under the same FDA submission

3 devices cleared under K083665, K854631, K854902 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.