Skip to content

ABX PENTRA Urine Control L/H recallZ-0924-2014 · Horiba Instruments, Inc dba Horiba Medical · initiated Dec 18, 2013

Class III Terminated All recalls posted in February 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Horiba Instruments, Inc dba Horiba Medical or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

HORIBA Medical is informing all ABX PENTRA Urine Control L/H customers running the Creatinine assay on ABX PENTRA Chemistry Analyzers that the Target Range for Creatinine has been updated. The ABX PENTRA Urine Control L/H lot specific annexes have been updated to include revised acceptable Target Ranges for Creatinine.No other analyte values are affected.

FDA root cause category: Pending

What the company is doing

HORIBA Medical sent an Urgent Safety Notice dated December 18, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to update their Quality Control procedures to include the new Creatinine Target Ranges for the ABX PENTRA Urine Control L/H Reference #: A11A01674 and Lot #s: 1207001, 1303601, and 1300701. Customers were instructed to complete and return the enclosed response form to verify their understanding of the Field Safety Notice. Customers with questions were instructed to contact their local HORIBA Medical representative. For questions regarding this recall call 949-453-0500.

Recall details

Recall number
Z-0924-2014
Recall class
Class III
Status
Terminated
Initiated by the company
Dec 18, 2013
Posted by the FDA
Feb 3, 2014
Recalling firm
Horiba Instruments, Inc dba Horiba Medical
Quantity
531
Product code
JJY
FDA submissions
K110137, K110530
Distribution
Worldwide Distribution - USA (nationwide) and Canada.

Product description

ABX PENTRA Urine Control L/H. The ABX PENTRA Urine Control L/H is a two-level (Low and High) quality control consisting of liquid solutions prepared from human urine with chemical additives and materials of biological origin added as required to obtain given component levels. The assigned values of the control components are given in the labeling, ensuring control of the appropriate HORIBA ABX SAS methods on the ABX PENTRA chemistry analyzers. Each control level is provided in one vial of 10 ml. The ABX PENTRA Urine Control L/H is for use in quality control by monitoring accuracy and precision.

Code information (models, lots, serial numbers)

Model/Part #(s): A11A01674. Lot/Serial #(s): 1207001, 1303601, and 1300701. Shelf Life: Lot#s 1207001 & 1300701 Expire on: 06/30/2014. Lot# 1303601 Expires on: 09/30/2014.

Other devices under the same FDA submission

1 device cleared under K110137, K110530 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page