Puritan Bennett 700 Series Ventilators (PB740 and PB760)… recallZ-0822-2014 · Nellcor Puritan Bennett Inc. (dba Covidien LP) · initiated Jan 25, 2013
Class II Terminated All recalls posted in January 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Nellcor Puritan Bennett Inc. (dba Covidien LP) or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Covidien is correcting labeling to clarify the operational life of the oxygen sensor (O2 sensor), PN 4-072214-00 used in the Puritan Bennett" 840, and PN, G-062010-00, used in Puritan Bennett" 740, and760 ventilator systems. The Operators Manual for PB840 states a nominal life of one year and the Addendum incorrectly states 2 years.
FDA root cause category: Labeling Change Control
What the company is doing
The firm, Covidien, sent an "URGENT MEDICAL DEVICE VOLUNTARY FIELD CORRECTION" letter dated January 24, 2014 to its Consignees/customers via certified mail. The letter described the product, problem and actions to be taken. The customers were instructed to read and understand the enclosed information; communicate this information to those who need to be aware of this labeling update; for the Puritan Bennett 840 ventilator only: If you have addendum P/N 066009A 09/02, remove and discard the addendum and replace it with the new information in Attachment I; for the Puritan Bennett 740 and 760 ventilators only: Add the new information in Attachment I; Ensure all O2 sensors in use or in inventory conform to the instructions in this notification. to your current operators manual and service manual, and read and understand the warnings, cautions and notes related to safety, ventilator use and operation that are contained in the Preface of the Operators Manuals. are asked to remove incorrect labeling and replace with the revised labeling. Covidien enclosed a labeling update (Attachment I) which provides detailed information related to the life of oxygen sensors used with the Puritan Bennett 840, 740 and 760 ventilators. Covidien has validated another oxygen sensor, P/N 10097559, for use in the Puritan Bennett 840, 740 and 760 ventilators. Information about that sensor is also included in the update in Attachment I. Should you have any questions regarding this letter or to report any issues with the Puritan Bennett" 840, 740 or 760 ventilators contact our Technical Support Department @ 1-800-255-6774, option 4, then option 1, to ensure proper device reporting procedures are followed.
Recall details
- Recall number
- Z-0822-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jan 25, 2013
- Posted by the FDA
- Jan 24, 2014
- Recalling firm
- Nellcor Puritan Bennett Inc. (dba Covidien LP)
- Quantity
- 7,959
- Product code
- CBK
- FDA submissions
- K964540, K970460, K984379, K990897
- Distribution
- Worldwide Distribution: US (nationwide) and Internationally to: Albania, Argentina, Australia, Bahrain, Belgium, Belize, Bermuda Bolivia, Bosnia and Herzegovina, Brazil, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Croatia, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Georgia, Germany, Greece, Guam, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macedonia, The former Yugoslav, Malaysia, Mexico, Moldova, Republic of, Montenegro, Morocco, Netherlands, Nicaragua, Northern Mariana Islands, Oman, Panama, Paraguay, Peru, Philippines, Poland, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Switzerland, Syria, Taiwan, Thailand, Tunisia, Turkey, Uganda, Ukrainian SSR, United Arab Emirates (UAE) United Kingdom, Uruguay, Venezuela and Yugoslavia.
Product description
Puritan Bennett 700 Series Ventilators (PB740 and PB760) PB700 Series Operators Manual: PN 10066984 & PN G-061874-00 and PB700 Series Service Manual: PN 10070389 The intended use of the oxygen sensor is to provide the oxygen concentration of the delivered gas at the patient wye (inspiratory-expiratory interface). The oxygen sensor does not control the oxygen level or the flow of gases. The intended use of the devices is as critical care ventilators that provide continuous mechanical ventilation for patients.
Code information (models, lots, serial numbers)
PB700 Series Operators Manual: PN 10066984 & PN G-061874-00 and PB700 Series Service Manual: PN 10070389
Other devices under the same FDA submission
1 device cleared under K964540, K970460, K984379, K990897 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Puritan Bennett G-061949-00Covidien LP · VENTILATOR, CONTINUOUS, FACILITY USE18 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 67163).
- Z-0821-2014 · TerminatedPuritan Bennett 840 Series Ventilator PB840 Operators Manual Addendum PN 066009A 09/02 The intended use of the oxygen sensor is to provide the oxygen concentration of the delivere…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page