7" (18 cm) Appx 0.33 ml, Smallbore Ext Set w/NanoClave"… recallZ-0694-2014 · ICU Medical, Inc. · initiated Dec 24, 2013
Class II Terminated All recalls posted in January 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from ICU Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
ICU Medical is recalling certain lots of the NanoClave Needlefree Connector product line because of a small percentage of devices that may leak.
FDA root cause category: Equipment maintenance
What the company is doing
icumedical sent an Urgent Medical Device Recall letter dated December 24, 2013, to all affected customers. The letter informed the customers of the problems identified and the actions to be taken. Customers were instructed to return completed verification forms. Customers with questions were instructed to contact ICU Customer Service Monday through Friday between the hours of 8:30AM and 4:00PM Pacific Time: (866) 829-9025 and select option 8, or email: [email protected]. ICU Medical reported an update on January 6, 2014, an expansion of the recall to additional lots. Updated recall letter was sent out to all customers on additional lots on January 6, 2014 via UPS with tracking. For questions regarding this recall call 949-366-2183.
Recall details
- Recall number
- Z-0694-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 24, 2013
- Posted by the FDA
- Jan 9, 2014
- Recalling firm
- ICU Medical, Inc.
- Quantity
- 4000 units
- Product code
- FPA
- FDA submissions
- K970855
- Distribution
- Worldwide Distribution - USA including UT, TX, GA, WI, VT, NH, AL, OH, TN, IL, CA, KY,HI, NC, and Internationally to CA, BE, IN, KW, and AU.
Product description
7" (18 cm) Appx 0.33 ml, Smallbore Ext Set w/NanoClave", Clamp, Rotating Luer, Part No. A1000, Item No. K7103-001 The ICU Needleless Connector is a single use, sterile, non-pyrogenic device intended for use as an accessory to intravascular administration set for the administration of fluids to a patient through a cannula placed in the vein or artery.
Code information (models, lots, serial numbers)
Lot No. 2767372
Other devices under the same FDA submission
11 devices cleared under K970855 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ICU Medical 011-MC100Icu Medical, Inc. · Set, administration, intravascular1 death93 reports
- ICU Medical B3300Icu Medical, Inc. · Set, administration, intravascular68 reports
- ICU Medical 011-C3300Icu Medical, Inc. · Set, administration, intravascular31 reports
- ICU Medical C1000Icu Medical, Inc. · Set, administration, intravascular22 reports
- ICU Medical 12568Icu Medical, Inc. · Set, administration, intravascular16 reports
- ICU Medical 11956Icu Medical, Inc. · Set, administration, intravascular12 reports
- ICU Medical 12512-01Icu Medical, Inc. · Set, administration, intravascular6 reports
- ICU Medical 011-C1000Icu Medical, Inc. · Set, administration, intravascular5 reports
- ICU Medical 011-A1000Icu Medical, Inc. · Set, administration, intravascular3 reports
- ICU Medical MC100-GRIcu Medical, Inc. · Set, administration, intravascular1 report
- ICU Medical SC9020Icu Medical, Inc. · Set, administration, intravascular1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 67166).
- Z-0658-2014 · TerminatedNanoClave Connector, Part No. A1000, Item No. 011-A1000, 061-A1000, A1000, K7083-001, Z3769 The ICU Needleless Connector is a single use, sterile, non-pyrogenic device intended fo…
- Z-0659-2014 · Terminated6" (15 cm) Appx 0.26 ml, Smallbore Ext Set with NanoClave T-Connector, Clamp, Luer Slip, Part No. A1000, Item No. 011-A1002 The ICU Needleless Connector is a single use, sterile, n…
- Z-0660-2014 · Terminated6" (15 cm) Appx 0.58 ml, Smallbore Trifuse Ext Set w/3 NanoClave, 3 Clamps, Rotating Luer, Part No. A1000, Item No. 011-A1029 The ICU Needleless Connector is a single use, sterile,…
- Z-0661-2014 · Terminated183 cm (72") Smallbore Quadfuse Ext Set w/4 NanoClave (Yellow, Green Rings), 4 Anti-Syphon Valves, 0.2 Micron Positive Filter, BCVClave , NanoClave T-Connector, Rotating Luer, Part…
- Z-0662-2014 · Terminated184 cm (72") Smallbore Bifuse Ext Set w/2 NanoClave, 2 Anti-Syphon Valves, 0.2 Micron Positive Filter, BCV-Clave, NanoClave TConnector, Rotating Luer, Part No. A1000, Item No. 011-…
- Z-0663-2014 · Terminated31 cm (12") Smallbore Bifuse Ext Set w/2 NanoClave, Orange Ring, 2 Anti Syphon Valves, 0.2 Micron Postive Filter, Clamp, NanoClave TConnector, Rotating Luer, Part No. A1000, Item N…
- Z-0664-2014 · TerminatedNanoClave Stopcock, Part No. A1000, Item No. 011-AC100 The ICU Needleless Connector is a single use, sterile, non-pyrogenic device intended for use as an accessory to intravascular…
- Z-0665-2014 · Terminated17 cm (7'') Ext Set w/NanoClave, 2 Gang 4-Way NanoClave Stopcock, Rotating Luer, Part No. A1000, Item No. 011-AC103 The ICU Needleless Connector is a single use, sterile, non-pyrog…
- Z-0666-2014 · Terminated6" (15 cm) Appx 0.26 ml, Smallbore Ext Set w/Clamp, NanoClave TConnector, Rotating Luer, Part No. A1000, Item No. A1001 The ICU Needleless Connector is a single use, sterile, non-p…
- Z-0667-2014 · Terminated7" (18 cm) Appx 0.29 ml, Smallbore Ext Set w/MicroClave Clear, NanoClave, Clamp, Rotating Luer, Part No. A1000, Item No. A1007 The ICU Needleless Connector is a single use, sterile…
- Z-0668-2014 · Terminated7" (18cm) Appx 0.32 ml, Smallbore Ext Set w/Remv MicroCLAVE Clear, NanoClave T-Connector, Clamp, Rotating Luer, Part No. A1000, Item No. A1009 The ICU Needleless Connector is a sin…
- Z-0669-2014 · Terminated8" (20 cm) Appx 0.51 ml, Smallbore Ext Set w/MicroCLAVE Clear, 0.2 Micron Filter, NanoClave T-Connector, Clamp, Rotating Luer, Part No. A1000, Item No. A1014 The ICU Needleless Con…
- Z-0670-2014 · Terminated78" Ext Set w/Check Valve, CLAVE Clear, Remv 2 Gang 4-Way Stopcocks, NanoClave, NanoClave T-Connector, 4 Clamps, Rotating Luer, 3 Ext, Part No. A1000, Item No. A1026 The ICU Needle…
- Z-0671-2014 · Terminated11" Smallbore Bifuse Ext Set w/1.2 Micron Filter, NanoClave", 3 Clamps, Rotating Luer, Part No. A1000, Item No. A1052 The ICU Needleless Connector is a single use, sterile, non-pyr…
- Z-0672-2014 · Terminated8" (20 cm) Appx 0.49 ml, Smallbore Bifuse Ext Set w/2 NanoClave", 2 Clamps, Luer Lock, Part No. A1000, Item No. A1069 The ICU Needleless Connector is a single use, sterile, non-pyr…
- Z-0673-2014 · Terminated8" Smallbore Pentafuse Ext Set w/5 NanoClave", 5 Clamps, Rotating Luer, Part No. A1000, Item No. A1070 The ICU Needleless Connector is a single use, sterile, non-pyrogenic device i…
- Z-0674-2014 · Terminated10" Smallbore Pentafuse Ext Set w/5 NanoClave", 5 Clamps, Rotating Luer, Part No. A1000, Item No. A1071 The ICU Needleless Connector is a single use, sterile, non-pyrogenic device …
- Z-0675-2014 · Terminated87" (221 cm) Appx 2.9 ml, Smallbore Trifuse Ext Set w/NanoClave, 0.2 Micron Filter, 4 Clamps, Rotating Luer, 2 Exts, Drop-In NanoClave, Part No. A1000, Item No. A1078 The ICU Needl…
- Z-0676-2014 · Terminated13" Smallbore Ext Set w/4 Gang 4-Way Stopcock w/Baseplate, 4 NanoClave, Rotating Luer, Part No. A1000, Item No. A1080 The ICU Needleless Connector is a single use, sterile, non-py…
- Z-0677-2014 · Terminated14" (36 cm ) Appx 0.50 ml, Smallbore Ext Set w/MicroClave Clear, NanoClave, 2 Clamps, Rotating Luer, Part No. A1000, Item No. A1081 The ICU Needleless Connector is a single use, s…
- Z-0678-2014 · Terminated103" (262 cm) Appx 3.9 ml, Smallbore Bifuse Ext Set w/NanoClave", Anti-Siphon Valve, Check Valve, 2 Clamps, Luer Lock, Part No. A1000, Item No. A1093 The ICU Needleless Connector …
- Z-0679-2014 · Terminated8" Ext Set w/NanoClave" Stopcock, Clamp, Rotating Luer, Part No. A1000, Item No. AC102 The ICU Needleless Connector is a single use, sterile, non-pyrogenic device intended for use…
- Z-0680-2014 · Terminated6" (15 cm) Appx 0.27 ml, Smallbore Ext Set w/Clamp, NanoClave" TConnector, Rotating Luer, Part No. A1000, Item No. K7084-001 The ICU Needleless Connector is a single use, sterile,…
- Z-0681-2014 · Terminated6" (15 cm) Appx 0.27 ml, Smallbore Ext Set w/Clamp, NanoClave" TConnector w/Luer Slip, Part No. A1000, Item No. K7085-001 The ICU Needleless Connector is a single use, sterile, no…
- Z-0682-2014 · Terminated7" (18 cm) Appx 0.29 ml, Smallbore Ext Set w/MicroClave Clear, NanoClave", Clamp, Rotating Luer, Part No. A1000, Item No. K7088-001 The ICU Needleless Connector is a single use, s…
- Z-0683-2014 · Terminated7" (18 cm) Appx 0.26 ml, Smallbore Ext Set w/MicroClave Clear, NThe ICU Needleless Connector is a single use, sterile, non-pyrogenic device intended for use as an accessory to intr…
- Z-0684-2014 · Terminated7" (18 cm) Appx 0.31 ml, Smallbore Ext Set w/Remv MicroClave Clear, NanoClave" T-Connector, Clamp, Rotating Luer, Part No. A1000, Item No. K7090-001 The ICU Needleless Connector i…
- Z-0685-2014 · Terminated10" (25 cm) Appx 2.9 ml, Trifuse Ext Set w/3 MicroClave Clear, Tri- Connector, NanoClave" T-Connector, 4 Clamps (Red, White, Blue), Rotating Luer, Part No. A1000, Item No. K7091-0…
- Z-0686-2014 · Terminated106" (269 cm) Appx 7.8 ml, Transfer Set w/Check Valve, NanoClave" TConnector, Anti-Siphon Valve, 2 Clamps, Luer Lock, Part No. A1000, Item No. K7093-001 The ICU Needleless Connect…
- Z-0687-2014 · Terminated78" Ext Set w/2 BCV-Clave, Remv 2-Gang 4-Way Stopcocks, NanoClave" T-Connector, 4 Clamps, Rotating Luer, 3 Exts, Part No. A1000, Item No. K7094-001 The ICU Needleless Connector is…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page