GE Healthcare Innova IGS 630. Biplane Cardiovascular and… recallZ-0853-2014 · GE Healthcare, LLC · initiated Nov 15, 2013

Class II Terminated All recalls posted in January 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

GE Healthcare has recently become aware of a potential safety issue with respect to IGS 630 Imaging Systems. While performing fluoroscopy on a biplane fluoroscopy unit, there is a potential for loss of the x-ray imaging function when the user changes the size of the lateral FOV (Field of View) and releases the biplane footswitch pedal simultaneously. This could lead to a system lock up requiring

FDA root cause category: Software design

What the company is doing

GE Healthcare sent an "Urgent Medical Device Correction" letter dated November 15. 2013. The letter was addressed to Hospital Administrators / Risk Mangers, Mangers of Radiology / Cardiology / Surgery Department and Radiologists / Cardiologists / Surgeons. The letter described the Safety Issue, Safety Instructions, Affected Product Details, and Product Correction. The letter stated that GE Healthcare will correct all affected systems at no cost. For questions call (In US 800-437-1171; for other countries contact your local GE Healthcare Service Representative).

Recall details

Recall number
Z-0853-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Nov 15, 2013
Posted by the FDA
Jan 28, 2014
Recalling firm
GE Healthcare, LLC
Quantity
28 (12 US, 16 OUS).
Product code
IZI
FDA submissions
K122457
Distribution
Worldwide Distribution - US Nationwide in the states of CA, CO, FL, NV, NJ, PA, TX and the countries of CZECH REPUBLIC, FRANCE, JAPAN, REPUBLIC OF KOREA, SAUDI ARABIA, and TURKEY.

Product description

GE Healthcare Innova IGS 630. Biplane Cardiovascular and Interventional Imaging System. Product Usage: The angiographic X-ray systems are indicated for use for patients from newborn to geriatric in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures. Additionally, with the OR table, the angiographic X-ray systems are indicated for use in generating fluoroscopic and rotational images of human anatomy for image-guided surgical procedures. The OR table is suitable for interventional and surgical procedures. The Cathiab Frontiers solutions are indicated for use in conjunction with single plane and biplane GE angiographic X-ray systems and imaging / data medical devices used in interventional and surgical cathlab environments and cleared for commercial distribution. The Cathiab Frontiers solutions are integrated GE angiographic X-ray and imaging / data medical devices that simplify the end-to-end clinical workflow in the cathlab by implementing: 1) Communication protocols for exchanging and automatically synchronizing patient, exam, system, and image information between the angiographic X-ray systems and the imaging / data medical devices. 2) Communication protocols for the control of imaging / data medical device functions from the angiographic X-ray systems user interface. 3) Interfaces for displaying the imaging / data medical device output on the monitor display solutions of the GE angiographic X-ray systems.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Serial # System ID 00000616724BU1 956389IGS630 00000618156BU4 MCMIGS630 FMI12200DUMMY2 702731ANGIO2 00000619372BU6 732828IGS30 00000618158BU0 210705CATH1 00000622076BU9 661327IGS630 00000621314BU4 904202WCL2 00000628656BU1 281MHWCL1 00000622220BU2 303839CL4 00000625357BU9 281420LAB3BP 00000626085BU5 817BGIGS630 00000628654BU6 215576IGS4 FMI12200DUMMY3 CS1340VA01 00000614683BU1 M4152252 00000618160BU6 YV1900 00000617143BU3 YV1901 00000618157BU2 YV1902 00000620495BU2 YV1903 00000624842BU1 YV1904 00000627935BU0 YV1905 00000630300BU2 YV1906 00000617142BU5 YV1950 FMI12200DUMMY4 YV1907 00000622672BU4 850060746 00000626239BU8 850060749 FMI12200DUMMY1 SA2308VA01 00000622221BU0 ZA2239VA01 00000614682BU3 34006VAS02

Other devices under the same FDA submission

1 device cleared under K122457 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.