ABX PENTRA Multical recallZ-1146-2014 · Horiba Instruments, Inc dba Horiba Medical · initiated Jan 20, 2014

Class II Terminated All recalls posted in March 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Horiba Instruments, Inc dba Horiba Medical or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

HORIBA Medical is informing all ABX PENTRA MultiCal, N Control, and P Control customers running the Creatinine Rate Blank Method assay on the ABX PENTRA 400 and PC200 that the Target Values for Creatinine Rate Blank Method have been updated. This issue is only applicable to Creatinine Rate Blank Method values reported on the ABX PENTRA 400 and PC200 Chemistry Analyzers only. No other analyzers

FDA root cause category: Under Investigation by firm

What the company is doing

HORIBA Medical initiated this recall by sending customers recall communication via FedEx week of January 20, 2014. The recall communication was a Field Safety Notice letter and Acknowledgement Form. The letter, dated 01/20/2014, titled "URGENT SAFETY NOTICE", informed customers of the recall by providing product description and codes, issue, impact, actiona/resolution, and contact information. The letter contained the following attachments: Field Safety Notice Acknowledgement Form and Updated Annexes with re-assigned values.

Recall details

Recall number
Z-1146-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Jan 20, 2014
Posted by the FDA
Mar 5, 2014
Recalling firm
Horiba Instruments, Inc dba Horiba Medical
Quantity
2,023 units
Product code
JIX
FDA submissions
K110530
Distribution
United States nationwide distribution.

Product description

ABX PENTRA Multical. ABX PENTRA N Control. ABX PENTRA P Control. The ABX PENTRA 400 and PC200 are benchtop clinical chemistry analyzers using two measuring principals: absorbance and ion selective electrodes. The ABX PENTRA Multical is a lyophilized human serum calibrator with chemical additives and materials of biological origin. The assigned values of the calibrator components are given in the enclosed annex, ensuring optimal calibration of the appropriate HORIBA ABX SAS methods on the ABX PENTRA Clinical Chemistry Analyzer. This calibrator is provided in ten vials of 3 ml.

Code information (models, lots, serial numbers)

Model Numbers:Multical - A11A01652, N Control - A11A01653, P Control - A11A01654.  Lot Numbers: MultiCal: 1204101 and 1304901, N Control: 1300801, P Control: 1300601.  Expiration Date (s): MultiCal: 1204101 Expires on 05/31/14, MultiCal: 1304901 Expires on 07/15/15, N Control: 1300801 Expires on 11/05/14, P Control: 1300601 Expires on 11/05/14.

Other devices under the same FDA submission

1 device cleared under K110530 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.