Reflection Interfit Shell, 52 MM OD, 3 HOLE, REF 71336052… recallZ-1066-2014 · Smith & Nephew Inc · initiated Jan 23, 2014
Class II Terminated All recalls posted in February 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smith & Nephew Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Inner spherical radius of the shell is undersized.
FDA root cause category: Under Investigation by firm
What the company is doing
Smith & Nephew Inc. sent an Urgent - Product Recall 1st Notification letter dated January 23, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. Actions for Hospital Representatives: 1. Please inspect your inventory and locate any unused devices from the above listed product and batch numbers, and quarantine them immediately. 2. Complete the last two columns in the Inventory Return Certification Form on the following page, indicating the quantities that need to be returned, and include your phone and fax number in the spaces provided. If you do not have product to return, please place an x in the column No product to return. 3. Please contact Smith & Nephews Safety Affairs Department via email at [email protected] or fax +1-901-566-7975 to obtain a return authorization (RA) number. 4. Return any affected product to the address listed on the Inventory Return Certification Form. Please indicate the RA number on your return shipment. Actions for Smith & Nephew Inventory Specialists 1. Please inspect your inventory and locate any unused devices from the above listed product and batch numbers, cease distribution and use of those products, and quarantine them immediately. 2. Carry out a physical count of all affected product in your territory and record this data on the Inventory Return Certification Form on the following page. 3. If you do not have product to return, please place an x in the column No Product to Return, and then fill in the Acknowledgement of Responsibility at the bottom of the form. 4. Fax a copy of the completed Inventory Return Certification Form to: +1-901-566-7975 5. If you have affected products, return the product, along with the completed Inventory Return Certification Form, to the address indicated
Recall details
- Recall number
- Z-1066-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jan 23, 2014
- Posted by the FDA
- Feb 21, 2014
- Recalling firm
- Smith & Nephew Inc
- Quantity
- 10 units
- Product code
- MBL
- FDA submissions
- K960094
- Distribution
- Worldwide Distribution: US Distribution to NJ and FL; and the countries of: Colombia and Costa Rica.
Product description
Reflection Interfit Shell, 52 MM OD, 3 HOLE, REF 71336052, QTY 1, STERILE R, Smith & Nephew, Inc. Indications for use include cemented or uncemented total hip arthroplasty for rheumatoid arthritis; osteoarthritis; post traumatic arthritis, avascular necrosis; femoral and neck fractures; osteomyelitis; fracture, dislocation of the hip; and unsuccessful cup arthroplasty, endoprosthesis, femoral osteotomy, or girdlestone resection.
Code information (models, lots, serial numbers)
Batch No. 13EM03053
Other devices under the same FDA submission
16 devices cleared under K960094 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- REFLECTION 71330001Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous20 reports
- REFLECTION 71336452Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous11 reports
- REFLECTION 71336454Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous9 reports
- REFLECTION 71336450Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous4 reports
- REFLECTION 71336456Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous4 reports
- REFLECTION 71336458Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous4 reports
- REFLECTION 71336054Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous3 reports
- REFLECTION 71336448Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous3 reports
- REFLECTION 71334056Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous1 report
- REFLECTION 71336046Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous1 report
- REFLECTION 71336460Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous1 report
- REFLECTION 71336462Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, porous1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page