CSI, Diamondback 360 Coronary Orbital Atherectomy System… recallZ-1095-2014 · Cardiovascular Systems, Inc. · initiated Jan 27, 2014
Class III Terminated All recalls posted in February 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cardiovascular Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Cardio vascular Systems Inc. has initiated a recall to request the immediate removal and return of Diamondback 360 Coronary Orbital Atherectomy Device (OAD). The device mistakenly contains a saline line that is used in our peripheral atherectomy device. This saline line was not part of the FDA approval for use in the coronary device. It therefore does not meet specification for this device. C
FDA root cause category: Process control
What the company is doing
Cardiovascular Systems Inc. sent an "Urgent Medical Device Recall" letter dated January 27, 2014 to all affected customers. The letter described the affected product, recall description and actions to be taken. Customers were instructed to complete and return the attached Customer Acknowledgement Form and return the device i available. For further information they should contact CSI Customer Service at 651 Campus Drive, Saint Paul, MN 55112, Tele 877-274-0901, Fax 612-677-3355.
Recall details
- Recall number
- Z-1095-2014
- Recall class
- Class III
- Status
- Terminated
- Initiated by the company
- Jan 27, 2014
- Posted by the FDA
- Feb 26, 2014
- Recalling firm
- Cardiovascular Systems, Inc.
- Quantity
- 318
- Product code
- MCX
- FDA submissions
- K071073
- Distribution
- USA Nationwide Distribution in the states of AZ, CA, FL, MA, MN, MS, NC, NJ, NY, PA, TN, TX, VA.
Product description
CSI, Diamondback 360 Coronary Orbital Atherectomy System, Part Number 70058-02, Model Number DBEC-125 Product Usage: The system is indicated to facilitate stent delivery in patients with coronary artery disease (CAD) who are acceptable candidates for PTCA or stenting due to de novo, severely calcified coronary artery lesions.
Code information (models, lots, serial numbers)
Lot: 85408 -85412, 85585 -85587, 87276, 87278, 87280, 87317, 87462, 87464, 87793, 87949 - 87952, 88520 - 88524, 88975 - 88979, 89498, 89501- 89504, 89952, 89954, 89956, 91876, 91877, 91891, 92124, 92156.
Other devices under the same FDA submission
1 device cleared under K071073 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- N/A Patient Data ModuleGE Medical Systems Information Technologies, Inc. · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page