Draeger PS500 Power Supply Unit of the Infinity ACS… recallZ-1298-2014 · Draeger Medical Systems, Inc. · initiated Feb 7, 2014

Class I Terminated All recalls posted in March 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Draeger Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The battery capacity of the optional PS500 Power Supply Unit of the Infinity ACS Workstation Critical Care (Evita Infinity V500) did not last as long as expected.

FDA root cause category: Component design/selection

What the company is doing

Drager sent an Urgent Medical Device Recall letter dated February 2014. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to use an affected device for patient transport only if absolutely necessary; not to rely on the battery charge status indicator, always supervise the patient and the ventilator during transport; ensure that they have a manual breathing bag available for manual ventilation as recommended in the instructions for use. If the power fail alarm occurs during transport, immediately provide manual ventilation and connect ventilator to a wall power source to resume ventilation. DragerService representative will schedule the implementation of a permanent solution as quickly as possible. The letter should be distributed to users of the Evita V500 and /or Babylog VN500 Ventilators to make them aware of this device behavior and risk mitigation measures. Customers with questions were instructed to call 1-800-543-5047. For questions regarding the operation and/or servicing of their Draer ventilator contact Technical Support at 1-800-543-5047. For questions regarding this recall call 215-721-5400.

Recall details

Recall number
Z-1298-2014
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Feb 7, 2014
Posted by the FDA
Mar 28, 2014
Recalling firm
Draeger Medical Systems, Inc.
Quantity
1544
Product code
CBK
FDA submissions
K093632, K093633
Distribution
Nationwide Distribution

Product description

Draeger PS500 Power Supply Unit of the Infinity ACS Workstation Critical Care (Evita Infinity V500), Evita V500 and /or Babylog VN500 Ventilators, used for ventilation during short term transport of patients within a hospital.

Code information (models, lots, serial numbers)

all lots

Other devices under the same FDA submission

2 devices cleared under K093632, K093633 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.