REF 623421 RS-ROM Noiles Rotating Hinge, Femoral Rotating… recallZ-1208-2014 · DePuy Orthopaedics, Inc. · initiated Feb 24, 2014
Class II Terminated All recalls posted in March 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from DePuy Orthopaedics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
DePuy Orthopaedics has identified the potential for holes to develop in the inner and outer flexible pouches that form the sterile barrier for S-ROM Noiles Rotating Hinge Femur with Pin devices. The outer carton and shrink wrap are intact. A package redesign is underway to resolve this issue. S-ROM Noiles Rotating Hinge Femur with Pin devices will remain in US Distributors inventory during the pa
FDA root cause category: Packaging
What the company is doing
DePuy Orthopaedics sent an Urgent Information - Recall Notice dated February 24, 2014, to all affected customers. The notice identified the product, reason for the recall and potential risks associated with using the recalled product. The notice includes response cards for the US Distributors to verify they have read, understood, and are adhering to the directions given in the field communication and for hospitals where consignment inventory is removed a hospital letter is available, as needed. No hospital signatures are required. Distributors are asked to send completed distributor cards to DePuy. Product related questions should be addressed to DePuy Orthopaedics, .Sales Consultant. Clinical questions should be addressed to DePuy Scientific Information Office 1-888-554-2482; recall questions to 574-371-4917.
Recall details
- Recall number
- Z-1208-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Feb 24, 2014
- Posted by the FDA
- Mar 13, 2014
- Recalling firm
- DePuy Orthopaedics, Inc.
- Quantity
- 132
- Product code
- JWH
- FDA submissions
- K896048
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Canada, Australia, Austria, China, Czech Republic, Denmark, Dubai / UAE, Estonia, France, Germany, Hungary, Iberia, India, Ireland, Israel, Italy, Malaysia, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Russia, Singapore, Slovenia, Spain, South Africa, Sweden, Switzerland, Turkey, United Kingdom, and Vietnam.
Product description
REF 623421 RS-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Right , X-Small, Sterile R Qty 1 The S-ROM Hinged knee femoral component is used in revision surgeries when there is significant bone loss or ligament instability and additional support is needed. S-Rom femorals and inserts are brought into nearly every revision case as backup should the MCL become compromised during the procedure. If additional constraint is required the surgeon can use the S-ROM femoral and insert. S-ROM femorals are used in less than 10% of knee revision procedures, but are present almost all the time.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Both Right and Left 102935, 110571, 114271, 114274, 114277, 128477, 132946, 142857, 156158, 162498, 162499, 162500, 172191, 172194, 193664, 193668, 217297, 224704, 229189, 229192, 229195, 229196, 236454, 236463, 236466, 241379, 249713, 249714, 249715, 254742, 254743, 260549, 260552, 269715, 269717, 278454, 278456, 283005, 283006, 283007, 291037, 291038, 291039, 295391, 302819, 306516, 306517, 314828, 314831, 314832, 317759, 317761, 317765, 317769, 317770, 317772, 329335, 338170, 338171, 338173, 338176, 346442, 354609, 354614, 360644, 360645, 360646, 360647, 366437, 366439, 366440, 373257, 373258, 378654, 378655, 378657, 378660, 385029, 385031, 385033, 385034, 392028, 392029, 392032, 400543, 400544, 402441, 403009, 403012, 409586, 414761, 414762, 414763, 414764, 414765, 414766, 414767, 430493, 430494, 430495, 430496, 430498, 438592, 438593, 438594, 438595, 438596, 438597, 438598, 438599, 448539, 452452, 453170, 453171, 453172, 456749, 456751, 456752, 456753, 456754, 470494, 470496, 470503, 470504, D51B54, D51B64, D51B84, D51B94, DN5ET4, DY5EP4, E4EGG4, E5WCF4, E6CFK4, FD6G14, FE2DP4, FL8AM4
Other devices under the same FDA submission
9 devices cleared under K896048 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- S-ROM NOILES 62-3401RDepuy (Ireland) · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER23 reports
- S-ROM NOILES 62-3411LDepuy (Ireland) · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER15 reports
- S-ROM NOILES 62-3411RDepuy (Ireland) · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER9 reports
- S-ROM NOILES 62-1601Depuy Orthopaedics, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER2 reports
- S-ROM NOILES 62-1602Depuy Orthopaedics, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER1 report
- S-ROM NOILES 62-1604Depuy Orthopaedics, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER1 report
- S-ROM NOILES 62-1613Depuy Orthopaedics, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER1 report
- S-ROM NOILES 62-1621Depuy Orthopaedics, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER1 report
- S-ROM NOILES 62-1623Depuy Orthopaedics, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 67627).
- Z-1203-2014 · TerminatedREF 623401 L S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Left, medium, Sterile R Qty 1 The S-ROM Hinged knee femoral component is used in revision su…
- Z-1204-2014 · TerminatedREF 623401 R, S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Right, Medium, Sterile R Qty 1 The S-ROM Hinged knee femoral component is used in revision …
- Z-1205-2014 · TerminatedREF 623411 L, S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Left, Small, Sterile R Qty 1 The S-ROM Hinged knee femoral component is used in revision su…
- Z-1206-2014 · TerminatedREF 623411 R S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Right, Small, Sterile R Qty 1 The S-ROM Hinged knee femoral component is used in revision su…
- Z-1207-2014 · TerminatedREF 623421 L S-ROM Noiles Rotating Hinge, Femoral Rotating Hinge Cemented (with Pin), Left , X-Small, Sterile R Qty 1 The S-ROM Hinged knee femoral component is used in revision …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page