Skip to content

Spacelabs Healthcare G2 Clinical Access Mislabeled CDs for… recallZ-1273-2014 · Spacelabs Healthcare, Llc · initiated Feb 14, 2014

Class II Terminated All recalls posted in March 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Spacelabs Healthcare, Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Spacelabs has voluntarily recalled G2 Clinical Access Software CD because the programs installed on the Clinical Access CD are incorrect and it will prevent the customer from installing Clinical Access.

FDA root cause category: Labeling Change Control

What the company is doing

Spacelabs Healthcare began sending out the Medical Device Correction letter, dated 28 February 2014, to domestic customers on 28 February 2014 and to international customers by 7 March 2014. The letter identified the product, the product, the problem, and the action to be taken by the customer. Customers are also directed to call 1-800-522-7025 and select 2 for Technical Support. Customers are advised to dispose the CD, PN: 063-1829-11 Rev A previously supplied and replace it with the replacement CD (Rev B) provided.

Recall details

Recall number
Z-1273-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Feb 14, 2014
Posted by the FDA
Mar 25, 2014
Recalling firm
Spacelabs Healthcare, Llc
Quantity
36 units total: 17 units were distributed in the US and 19 units distributed outside the US.
Product code
DQK
FDA submissions
K110779
Distribution
Worldwide Distribution- USA (nationwide) including the states of Alabama, Florida, Georgia, Louisiana, Missouri, New York, South Carolina, Tennessee, Texas and Wyoming, and the countries of Australia, Canada, Czechoslovakia, China, Colombia, France, Great Britain, Italy, Netherlands, Poland, Saudi Arabia, and Turkey.

Product description

Spacelabs Healthcare G2 Clinical Access Mislabeled CDs for PN: 063-1829-11 Rev A and Update Kit, PN: 040-1565-10. Used for installing ICS G2 Clinical Access software onto hospital personal computers.

Code information (models, lots, serial numbers)

Two G2 Clinical Access CD part numbers are affected: PN: 063-1829-11 Rev A and Update Kit, PN: 040-1565-10.  Serial numbers distributed in the US: "040-1565-10" and "063-1829-11".   Serial number distributed outside the US: "063-1829-11".

Other devices under the same FDA submission

1 device cleared under K110779 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page