The OEC 9900 Elite mobile fluoroscopy system (image… recallZ-1462-2014 · GE OEC Medical Systems, Inc · initiated Mar 31, 2014
Class II Terminated All recalls posted in April 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE OEC Medical Systems, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE Healthcare has identified a potential safety issue related to the screws that attach the workstation monitor spring arm to the extension arm on certain OEC 9900 systems, causing the spring arm and monitor to fall forward.
FDA root cause category: Device Design
What the company is doing
GE Healthcare sent an Urgent Medical Device Correction letter dated March 31, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Use caution when positioning or extending the workstation articulating arm. In addition, refrain from positioning staff or patients directly underneath or in front of a workstation monitor. Your local GE Healthcare Field Engineer will contact you to coordinate the replacement of the screws, and ensure that your system is operating safely and meets all specifications. These activities will be performed at no cost to you. Should you have any questions or concerns regarding these corrections, please do not hesitate to contact the service team for further information at 800-874-7378 option 8. Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact us immediately. UPDATE: On June 20, 2014. GE Healthcare Surgery began distributing an Urgent Medical Device Correction Letter to customers with affected units. The letter makes clear that this field correction applies to all user sites with the listed products, regardless of whether service work was performed on the monitor extension arm since 3/31/2014.
Recall details
- Recall number
- Z-1462-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Mar 31, 2014
- Posted by the FDA
- Apr 16, 2014
- Recalling firm
- GE OEC Medical Systems, Inc
- Quantity
- 453 units
- Product code
- OWB
- FDA submissions
- K061953, K073543, K082781
- Distribution
- Worldwide Distribution - USA, and the countries of Australia, Argentina, Canada, China, Cambodia, Columbia, Japan, Ecuador, Japan, Korea, Hong Kong, India, Israel, Libya, Mexico, Nigeria, Peru, Russia, Thailand, Taiwan, Saudi Arabia and Singapore.
Product description
The OEC 9900 Elite mobile fluoroscopy system (image intensified fluoroscopic x-ray) is designed to provide fluoroscopic and spot-film images of the patient during diagnostic surgical and interventional procedure. Examples of clinical application may include cholangiography, endoscopy, urologic, orthopedic, neurologic, vascular, cardiac, critical care and emergency room procedure. The system may be used for other imaging applications at the physician's discretion.
Code information (models, lots, serial numbers)
Model Numbers - OEC 9900 Elite
Other devices under the same FDA submission
1 device cleared under K061953, K073543, K082781 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- OEC 9900 ELITEOec Medical Systems, Inc. · Interventional fluoroscopic x-ray system3 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗