Skip to content

IMMULITE 2000 /IMMULITE 2000 XPi Systems TIE Total IgE, REF… recallZ-1518-2014 · Siemens Healthcare Diagnostics · initiated Dec 20, 2013

Class II Terminated All recalls posted in April 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Siemens confirmed an overall average positive bias of 23% against the WHO 2nd IRP 75/502 with the IMMULITE Systems Total IgE assays.

FDA root cause category: Under Investigation by firm

What the company is doing

Urgent Medical Device Correction Letters (dated 12/20/13) and effectiveness check response forms were sent to customers on 12/20/13 via FedEx to inform them of the issue. Customers were requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Correction letter within thirty (30) days. Field service personnel were sent a support document describing the issue and instructing them how to deal with customer questions.

Recall details

Recall number
Z-1518-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Dec 20, 2013
Posted by the FDA
Apr 25, 2014
Recalling firm
Siemens Healthcare Diagnostics
Quantity
18,489 units
Product code
JHR
FDA submissions
K931703
Distribution
Worldwide distribution.

Product description

IMMULITE 2000 /IMMULITE 2000 XPi Systems TIE Total IgE, REF L2KIE2, SMN 10380873, and L2KIE6, SMN 10380872, IVD --- Siemens Healthcare Diagnostics Products Ltd. For in vitro diagnostic use with the IMMULITE, IMMULITE 1000, and IMMULITE 2000/ IMMULITE 2000 XPi Systems Analyzers - for the quantitative measurement of Total Immunoglobulin type E (IgE) in serum.

Code information (models, lots, serial numbers)

Lots 260 through 275

Other devices under the same FDA submission

1 device cleared under K931703 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 67210).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page