Pipeline Embolization Device (PED). Used endovascular… recallZ-1453-2014 · Micro Therapeutics Inc, Dba Ev3 Neurovascular · initiated Apr 1, 2014
Class I Terminated All recalls posted in April 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Micro Therapeutics Inc, Dba Ev3 Neurovascular or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for the PTFE (polytetrafluoroethylene) coating to delaminate and detach from the delivery wire.
FDA root cause category: Process change control
What the company is doing
A customer notification letter was sent on 4/1/14 to all customers who purchased the Pipeline Embolization Device and the Alligator Retrieval Device. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete the customer response form and fax the completed form to Covidien at (763) 591-3519. Customers with questions are instructed to contact Dr. Mark A. Turco at (202) 310-5120. A press release was sent on 4/11/14 to Business Wire Newsroom to inform the public about the recall from Covidien. The recall is regarding certain lots of its Pipeline Embolization Device and Alligator Retrieval Device where the polytetrafluoroethylene (PTFE) coating applied to the delivery wire could delaminate adn detach from the devices. Consumers are instructed to report any adverse reactions or quality problems experienced with the use of the product to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax. Consumers with may also contact Covidien Customer Service at (800) 716-6700 between 7am to 7pm (central) or email at [email protected].
Recall details
- Recall number
- Z-1453-2014
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Apr 1, 2014
- Posted by the FDA
- Apr 16, 2014
- Recalling firm
- Micro Therapeutics Inc, Dba Ev3 Neurovascular
- Quantity
- 32 units total (23 units in US)
- Product code
- DQY
- FDA submissions
- P100018
- Distribution
- Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Germany, France, UK, Turkey, Argentina, Canada, Colombia, Mexico, , Austria, Belgium, Denmark, Hungary, Italy, Netherlands, Norway, Poland, Portugal, Russia, Slovakia, Slovenia, Spain, Switzerland, and UAE.
Product description
Pipeline Embolization Device (PED). Used endovascular treatment of adults with intracranial aneurysms in the internal carotid artery.
Code information (models, lots, serial numbers)
PED Model/ Lot FA7735014 / 9879718 FA7742512 / 9864827 FA7150035 / 9855921 FA7735016 / 9856429 FA7742518 / 9856431 FA7730010 / 9869991 FA7737514 / 9855933 FA7742518 / 9860762 FA7732514 / 9873854 FA7737516 / 9858010 FA7745014 / 9874111 FA7735014 / 9855265 FA7740016 / 9859214 FA7745016 / 9879736 FA7735014 / 9875267 FA7740016 / 9879180 FA7747514 / 9868888
Other devices under the same FDA submission
292 devices cleared under P100018 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Pipeline TM Flex Embolization Device with Shield Technology TM PED2-300-12Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter2 recalls1 report
- PIPELINE PED-500-20Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall31 reports
- PIPELINE PED-500-35Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death29 reports
- PIPELINE PED-500-16Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall2 deaths27 reports
- PIPELINE PED-500-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death23 reports
- PIPELINE PED-500-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall21 reports
- PIPELINE PED-450-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall20 reports
- PIPELINE PED-475-30Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death19 reports
- PipelineTM Flex Embolization Device with Shield TechnologyTM PED2-500-12Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter67 reports
- PIPELINE PED-400-25Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall17 reports
- PIPELINE PED-475-14Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall16 reports
- PIPELINE PED-450-35Micro Therapeutics, Inc. · Intracranial aneurysm flow diverter1 recall1 death15 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page