HexaPOD evo RT Couchtop recallZ-1514-2014 · Elekta, Inc. · initiated Apr 17, 2014
Class II Terminated All recalls posted in April 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Elekta, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
If screws are not properly tightened with a torque wrench they may become loose over time.
FDA root cause category: Component design/selection
What the company is doing
Corrective Action #1 - Important Field Safety Notice 618 01 302 002 was issued April 2014. The Notice informs end users about the problem and the potential clinical impact. It also give a recommendation how the risk can be mitigated - inspect the screws and if loose contact Elekta Service to tighten the screws with 20 Nm. The notice also advises the customer that a modification kit will be released that introduces an improved screw solution. The solution will be implemented on all affected products when available. Corrective Action #2 Permanent Solution - Replace Current Screws - An improved screw has been designed and will be available for Elekta Field Service organization to replace on all affected products at the end of April. The new design solution provides a very long used thread depth, exceeding the standard recommended thread-depths by far. This design will prevent the screws from loosening.
Recall details
- Recall number
- Z-1514-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Apr 17, 2014
- Posted by the FDA
- Apr 25, 2014
- Recalling firm
- Elekta, Inc.
- Quantity
- 266
- Product code
- JAI
- FDA submissions
- K072898
- Distribution
- Worldwide Distribution-USA including DC and the states of AZ, CA, CT, FL, GA, IL, HI, KY, MD, MA, MI, MS, MO, NJ, NY, NH, NC, ND, OH, OK, OR, PA, SC, TX, VA, WA, WI, and WY, and the countries of Algeria, Australia, Austria, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kazakhstan, Mexico, Netherlands, New Zealand, Peru, Poland, Portugal, Russia, Saudi Arabia, South Korea, Span, Sweden, Switzerland, Taiwan, Thailand, Turkey, and United Kingdom.
Product description
HexaPOD evo RT Couchtop. To support and aid in positioning a patient during radiation therapy.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Serial Numbers 42188, 42246, 42106, 42153, 42107, 42068, 42012, 42013, 42062, 42009, 42224, 42126, 3005, 42222, 3019, 3000, 42058, 42059, 42057, 42187, 42129, 42159, 3022, 3017, 42189, 42213, 42202, 42212, 42004, 4220, 42113, 42178, 42216, 42207, 2982, 2986, 2991, 42102, 42000, 42032, 42145, 42174, 42008, 42048, 42064, 42070, 42074, 42084, 42115, 42134, 42118, 42152, 42137, 42146, 3008, 2987, 3010, 3018, 42069, 42036, 3029, 42162, 42139, 42011, 42007, 42173, 42109, 42193, 3006, 42053, 42015, 42105, 42124, 42085, 42065, 42094, 42198, 2998, 3003, 3033, 3024, 2980, 42121, 2999, 42027, 42061, 42075, 42079, 42214, 3020, 42192, 42247, 42202, 42155, 42025, 42021, 42127, 42151, 42239, 42033, 42035, 42089, 42204, 42200, 2995, 3030, 42266, 42197, 42077, 42001, 42037, 42147, 42186, 42016, 42028, 42029, 42050, 42042, 42054, 42047, 42051, 42055, 42049, 42262, 42080, 42076, 42063, 42103, 42087, 42117, 42083, 42166, 42101, 42227, 42150, 42156, 42163, 42179, 42180, 42140, 42165, 42157,42234, 42169, 42182, 42183, 42236, 42195, 42223, 42256, 42276, 42260, 42231, 42250, 42243, 42240, 42272, 42235, 42249, 42254, 42245, 42265, 42268, 42267, 42253, 42264, 2993, 2997, 3004, 3031, 3002, 42226, 42258, 42114, 42081, 42090, 42093, 42133, 42181, 42039, 42116, 3016, 29677, 42010, 42073, 42176, 42164, 42031, 3028, 42041, 42038, 42022, 42034, 42088, 42104, 2976, 2979, 2988, 3011, 3015, 3021, 3027, 42238, 42018, 42135, 42252, 42185, 42273, 42060, 42024, 42006, 42082, 42194, 42263, 42232, 42199, 42170, 42161, 42078, 42158, 2985, 2994, 42120, 3032, 42056, 2969, 3034, 2975, 2972, 2984, 42191, 42225, 42228, 42023, 42143, 42122, 42208, 42230, 42211, 42026, 42206, 42066, 42052, 42014, 42218, 42030, 42040, 42244, 42043, 42044, 42045, 42046, 42123, 42255, 42141, 42125, 42072, 42112, 42210, 42017, 420196, 42020, 42067, 42071, 42097, 42100, 42131, 42144, 42136, 42148, 42130, 42111, 42067, 42171, 42203, 4210, 42205, 42096, 42154, 42241, 42229, 2981, 2989, 2971, 2990, 3007, 3013, 2992, 2973, 2978, 30098, 2974, 3014, 3012, 3025, 2996, 3026, 42274
Other devices under the same FDA submission
1 device cleared under K072898 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- HexaPOD P10603-1003C-Medical Intelligence GmbH · Couch, Radiation Therapy, Powered1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.