VersaCell X3 Solution and VersaCell X3 Expansion Pack, a… recallZ-1832-2014 · Siemens Medical Solutions Diagnostics · initiated Apr 25, 2014
Class III Terminated All recalls posted in June 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Medical Solutions Diagnostics or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The keyboard and monitor may not be properly secured to the monitor arm.
FDA root cause category: Device Design
What the company is doing
Siemen's Healthcare Diagnostics sent an Urgent Medical Device Correction letter dated April 25, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to complete and return teh Field Correction Effectiveness Check form attached to the letter within 30 days by fax to 302-631-7597. Inspect the monitor, keyboard and bracket assembly. If the screw and washer assembly is secure as shown in Figure 1, the monitor and keyboard assembly is safe to use. After making adjustments or repositioning the monitor, keyboard and bracket assembly, inspect the screw adn washer to ensure it is ssecure as shown in Figure 1. If there is a gap between the underside of the monitor support bracket and cylindrical mounting feature as shown in Figure 2, contact your Siemens technical support representative to schedule a service visit as soon as possible. Front load and manage samples directly on the connected instruments until your Siemens technical support representative addresses the issue. Customers were instructed to retain the letter with their laboratory records, and forward the letter to those who may have received the affected product. Customers with questions were instructed to contact their Siemens Customer Care Center or their local Siemens technical support representative. For questions regarding this recall call 877-229-3711.
Recall details
- Recall number
- Z-1832-2014
- Recall class
- Class III
- Status
- Terminated
- Initiated by the company
- Apr 25, 2014
- Posted by the FDA
- Jun 19, 2014
- Recalling firm
- Siemens Medical Solutions Diagnostics
- Quantity
- 19 units
- Product code
- LXG
- FDA submissions
- K970227
- Distribution
- Worldwide Distribution - USA including Iowa and New York and Internationally to Canada, Chile, France, India, Italy, Malaysia, Peru, Korea, and the United Kingdom.
Product description
VersaCell X3 Solution and VersaCell X3 Expansion Pack, a robotic laboratory sample-management system. It connects up to 3 instruments and provides a user interface that routes samples, reduces manual tasks and consolidates results.
Code information (models, lots, serial numbers)
Siemens Material Numbers: Instrument 10793839, Expansion Pack 10913528
Other devices under the same FDA submission
3 devices cleared under K970227 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- IMMULITE® 2000 XPi Immunoassay System 10373214Siemens Healthcare Diagnostics Products Ltd · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE286 reports
- VersaCell® X3 10793839Siemens Healthcare Diagnostics Products Ltd · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE3 reports
- IMMULITE® 2000 Systems TSH Third Generation TSH 10381665Siemens Healthcare Diagnostics Products Ltd · RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page