Skip to content

VerSys Hip System, Beaded Fullcoat Stems, Zimmer, Inc.… recallZ-1842-2014 · Zimmer, Inc. · initiated May 19, 2014

Class II Terminated All recalls posted in June 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During routine inspection, beaded hip stem forgings received from supplier failed to meet ZES 2A-102. Processing of Beaded CoCrMo Alloy Hip Stem, material requirements per ATS #14-04818 and ATS #14-05549, respectively.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On 6/2/2014 the firm sent "Urgent Medical Device Recall" letters to their customers. On 8/13/14, Zimmer contacted a distributor via phone pertaining to one additional lot (00-7843-011-06; lot 62547211) which had been included in the recall. Formal recall notices were issued to the distributor via email on 8/15/14.

Recall details

Recall number
Z-1842-2014
Recall class
Class II
Status
Terminated
Initiated by the company
May 19, 2014
Posted by the FDA
Jun 20, 2014
Recalling firm
Zimmer, Inc.
Quantity
147 units
Product code
LPH
FDA submissions
K973714
Distribution
Worldwide Distribution-USA (nationwide) including the states of AL, AZ, CA, CO, CT, FL, GA, IL, IN, MA, MI, MN, NC, ND, NJ, NY, OH, OK, PA, TN, TX, VA & WI, and the countries of Canada, Colombia, Dominican Republic, Japan, Korea (S. Korea), Peru, Taiwan & Thailand.

Product description

VerSys Hip System, Beaded Fullcoat Stems, Zimmer, Inc., Warsaw, IN 46581

Code information (models, lots, serial numbers)

Item # 784301108, Lot #'s: 62478967, 62479742, 62503408, 62528282, 62530786, 62544915, 62554360; Item # 784301126, Lot #: 62544885, 62547211, Item # 784301156, Lot #, 62587079; Item # 784301208, Lot #'s: 62503411, 62509385, 62530790, 62547217, 62554362; Item # 784301226, Lot #'s: 62515502, 62547216; Item # 784301256, Lot #'s: 62489572, 62522386; Item # 784301326, Lot #, 62496464, 62594428, 62622893; Item # 784301356, Lot #: 62622894; Item # 784301382, Lot #: 62544907; Item # 784301511, Lot #: 62472423 & Item # 784301581, Lot #: 62496469.

Other devices under the same FDA submission

4 devices cleared under K973714 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page