Why it was recalled

There is a misassembly issue with the photometer installed on Dimension Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.

FDA root cause category: Employee error

What the company is doing

An urgent field safety notice, dated May 2014, was sent to users to inform them of the issue and that a service representative would replace the misassembled photometer.

Recall details

Recall number
Z-2133-2014
Recall class
Class III
Status
Terminated
Initiated by the company
May 27, 2014
Posted by the FDA
Jul 30, 2014
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
15
Product code
JJE
FDA submissions
K073604, K112999, K130276
Distribution
FL LA AR IA OH OH WI IL LA KS MT

Product description

Dimension Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.

Code information (models, lots, serial numbers)

DE271317, DE271320, DE271323, DE271324, DE271326, DE271327, DE271328, DE271332 DE271341, DE271343, DE271325, 12252589, 2004082898, 2004082899, 224719-AX

Other devices under the same FDA submission

2 devices cleared under K073604, K112999, K130276 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.