Skip to content

DISCOVERY ELBOW prosthesis, 4 MM X 100 MM RIGHT HUMERAL… recallZ-1925-2014 · Biomet, Inc. · initiated May 15, 2014

Class II Terminated All recalls posted in June 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Surface finish is different than specified. The implant has 30 grit blast on the exterior of the flange when the print calls for the glass bead blast. The patient may experience pain requiring surgical intervention.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Biomet sent an Urgent Medical Device Recall Notice letter dated May 28, 2014 to their customer. The letter identified the affected product, problem and actions to be taken. The letter instructed physician to monitor the patient for any pain that may require surgical intervention. For questions call (574) 372-1570.

Recall details

Recall number
Z-1925-2014
Recall class
Class II
Status
Terminated
Initiated by the company
May 15, 2014
Posted by the FDA
Jun 26, 2014
Recalling firm
Biomet, Inc.
Quantity
1
Product code
JDC
FDA submissions
K013042
Distribution
US Nationwide Distribution in the state of California

Product description

DISCOVERY ELBOW prosthesis, 4 MM X 100 MM RIGHT HUMERAL FLANGED I BOND COAT Tl 6AL 4V ALLOY. Elbow joint replacement prostheses are intended for primary and revision joint arthroplasty for use in cemented applications.

Code information (models, lots, serial numbers)

Catalog/REF #: 114905, Lot #: M846640

Other devices under the same FDA submission

12 devices cleared under K013042 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page