ADVIA Centaur Systems TnI-Ultra assay recallZ-1991-2014 · Siemens Healthcare Diagnostics, Inc · initiated May 30, 2014
Class II Terminated All recalls posted in July 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The solid phase reagent in some of the ADVIA Centaur TnI -Ultra ReadyPacks, lot 084, is darker in appearance, creating a potential for incorrect control and patient results.
FDA root cause category: Nonconforming Material/Component
What the company is doing
An Urgent medical Device Correction (UMDC) was mailed, via FedEx, to all affected Siemens Healthcare Diagnostics customers in the United States on May 30, 2014 for customer receipt on June 2, 2014. All international customers were provided with a copy of the Urgent Field Safety Notice via hard copy, e-mail, and/or fax as determined by each country's local regulations and procedures. Each customer is asked to complete a fax-back form. This notice informed the customers of the potential for incorrect control and patient results due to solid phase which is darker in appearance. The notice instructs customers to place one TnI-Ultra ReadyPack on the system at a time and to calibrate and run controls on each ReadyPack to ensure accurate control and patient values. Patient TnI-Ultra results produced with this assay are acceptable if they follow acceptable calibration with valid quality control results. Upon receipt of kit lots ending in 086 and higher, calibration of each ReadyPack is not required. This notice should be reviewed with the facility's laboratory or medical director.
Recall details
- Recall number
- Z-1991-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 30, 2014
- Posted by the FDA
- Jul 2, 2014
- Recalling firm
- Siemens Healthcare Diagnostics, Inc
- Quantity
- 34,283 kits
- Product code
- MMI
- FDA submissions
- K053020
- Distribution
- Worldwide Distribution-USA (nationwide) including PR and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY, and the countries of Canada, Mexico, Argentina, Austria, Australia. Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, French Polynesia, Great Britain, Greece, Guadeloupe, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Latvia, Lebanon, Libya, Macedonia, Malaysia, ,Malta, Martinique, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Republic of Korea, Romania, Russian Federation, Serbia, Saudi Arabia, Singapore, Slovakia, Slovenia, Spain, Sri Lanka, South Africa, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Uruguay, and Vietnam.
Product description
ADVIA Centaur Systems TnI-Ultra assay; Catalog Number 02789602 (100 tests), SMN 10317708 and 02790309 (500 tests), SMN 10317790
Code information (models, lots, serial numbers)
Kit Lots Ending In: 078, 079, 082, 083, 084
Other devices under the same FDA submission
4 devices cleared under K053020 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ADVIA Centaur® TnI-Ultra™ 10317708Siemens Healthcare Diagnostics Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT45 reports
- ADVIA Centaur® TnI-Ultra™ 10317709Siemens Healthcare Diagnostics Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT36 reports
- Atellica IM TnI-Ultra 10995696Siemens Healthcare Diagnostics Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT6 reports
- Atellica IM TnI-Ultra 10995697Siemens Healthcare Diagnostics Inc. · IMMUNOASSAY METHOD, TROPONIN SUBUNIT3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page