Zimmer EDI PERIARTICULAR PLATING SYSTEM CANCELLOUS BONE… recallZ-2032-2014 · Zimmer Manufacturing B.V. · initiated Jun 12, 2014
Class II Terminated All recalls posted in July 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Manufacturing B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Two lots of screws were commingled. Screws with etching and machining for lot number 62629432 were packaged in bags labeled for lot number 62628759; and screws with etching and machining for lot number 62628759 were packaged in bags labeled for lot number 62629432. The affected product was distributed between 3/6/14 and 4/21/14.
FDA root cause category: Process design
What the company is doing
Zimmer sent an Urgent Medical Device Recall Lot specific letter dated June 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed: 1. Please locate, quarantine, and return the affected product (part number 00-4840-020-00 lot number 62629432 and part number 00-2347-022-14 lot number 62628759) from all inventories. 2. Please note that any hospitals that received direct shipments of this product from Zimmer will be notified via Fed Ex. 3. Please review the list of hospital and surgeons to be provided to your territory via email. 4. Provide the name, address, and phone number of any additional hospitals or surgeons that may have received or used the affected product, or confirm that the list provided is complete. 5. Complete and return Attachment 1. 6. Ensure that a copy of the Risk Managers and Surgeons letter is left with the facilities from which product is removed by providing them with a copy of the letter, if they are not identified as a facility receiving a letter directly from Zimmer. 7. If after reviewing this notification you have questions or concerns, please contact Zimmer at 1-877-946-2761.
Recall details
- Recall number
- Z-2032-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jun 12, 2014
- Posted by the FDA
- Jul 10, 2014
- Recalling firm
- Zimmer Manufacturing B.V.
- Quantity
- 496
- Product code
- HWC
- FDA submissions
- K934765
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of India, Japan, China, Australia and Canada.
Product description
Zimmer EDI PERIARTICULAR PLATING SYSTEM CANCELLOUS BONE SCREW, FULLY THREADED, HEX HEAD 4MM DIAMETER, 14 MM LENGTH, and EDI 4.0 MM CANCELLOUS SCREW, FULLY THREADED, 20 MM LENGTH. Temporary internal fixation devices, designed to stabilize fractures during the normal healing process.
Code information (models, lots, serial numbers)
Lot No. 62629432 Lot No. 62628759
Other devices under the same FDA submission
8 devices cleared under K934765 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Trilogy® 6200-56-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED5 reports
- Trilogy® 6200-50-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 reports
- Trilogy® 6200-52-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- Trilogy® 6200-60-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- Trilogy® 6200-54-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Trilogy® 6200-62-20Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Trilogy® 6200-66-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Trilogy® 6105-50-32Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page