Nanosphere Verigene recallZ-2168-2014 · Nanosphere, Inc. · initiated Oct 16, 2013
Class III Terminated All recalls posted in August 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Nanosphere, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Nanosphere has received report of an increased rate of "No Call - INT CTL 2" results for the Internal Control 2 (INT CTL 2) when negative samples are tested with certain lots of CDF Amplification Trays (20-011-022), which are contained in the CDF Amplification Reagent Kit (20-012-022); specifically for lots 061913022D and 071813022E.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Nanosphere sent an Field Notification Letter dated October 16, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letters included instructions for customers to: 1) discontinue use of the affected product, 2) discard any of the affected product, and 3) complete the attached form and return it to Nanosphere. Customers with questions can contact Nanosphere at 847-400-9000 or 888-837-4436 x 2.
Recall details
- Recall number
- Z-2168-2014
- Recall class
- Class III
- Status
- Terminated
- Initiated by the company
- Oct 16, 2013
- Posted by the FDA
- Aug 8, 2014
- Recalling firm
- Nanosphere, Inc.
- Quantity
- 280 sets of trays
- Product code
- OZN
- FDA submissions
- K123197
- Distribution
- US Distribution including the states of CA and WA
Product description
Nanosphere Verigene Each CDF test consists of a Verigene CDF Nucleic Acid Test Kit and a Verigene CDF Amplification Reagent Kit. The Verigene Clostridium difficile Nucleic Acid Test (CDF) is a qualitative multiplexed in vitro diagnostic test for the rapid detection of toxin A (tcdA), toxin B (tcdB), and tcdC gene sequences of toxigenic Clostridium difficile and for presumptive identification of PCR ribotype 027 strains from unformed (liquid or soft) stool specimens collected from patients suspected of having C. difficile infection (CDI). Presumptive identification of the PCR ribotype 027 strain of C. difficile is by detection of the binary toxin (cdt) gene sequence and the single base pair deletion at nucleotide 117 in the tcdC gene. The tcdC gene encodes for a negative regulator in C. difficile toxin production. The test is performed on the Verigene System and utilizes automated specimen preparation and polymerase chain reaction (PCR) amplification, combined with a nanoparticle-based array hybridization assay to detect the toxin gene sequences associated with toxin-producing C. difficile. The CDF test is indicated for use as an aid in the diagnosis of CDI. Detection of PCR ribotype 027 strains of C. difficile by the CDF test is solely for epidemiological purposes and is not intended to guide or monitor treatment for C. difficile infections. Concomitant culture is necessary only if further typing or organism recovery is required.
Code information (models, lots, serial numbers)
Verigene CDF Amplification Trays, Catalog Number: 20-011-022, Lot Numbers: 061913022D and 071813022E
Other devices under the same FDA submission
1 device cleared under K123197 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Clostridium difficile (CDF) Stool Sample Preparation Kit 30-002-022Luminex Corporation · C. Difficile toxin gene amplification assay1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.