Verigene BC-GN Kit (Catalog number 20-005-021) recallZ-2736-2014 · Nanosphere, Inc. · initiated Jun 17, 2013

Class II Terminated All recalls posted in September 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Nanosphere, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a specific Extraction Tray lot containing Tips that may slightly increase the occurrence of Process SP Processing Error when used due to the Processor SP experiencing difficulty releasing the Tip to the final location in the Tip Assembly. The difficult releasing the Tip can result in a Processing Error and occasionally the Tip to be dropped into the Processor SP.

FDA root cause category: Process change control

What the company is doing

Undated recall notification letters were sent to all direct accounts (customers) June 17 - 19, 2013. The letters included instructions for customers to: 1) check their inventory for the recalled products; 2) set any recalled products aside for return; and, 3) contact Nanosphere to obtain replacement tip holders. Customers with questions about this recall can contact Nanosphere via e-mail at [email protected] or by calling 888-837-4436.

Recall details

Recall number
Z-2736-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 17, 2013
Posted by the FDA
Sep 30, 2014
Recalling firm
Nanosphere, Inc.
Quantity
20 extraction trays
Product code
PEN
FDA submissions
K132843
Distribution
Worldwide Distribution:US (Nationwide) including states of: AZ, CA, FL, IL, KY, MD, MN, MT, OH, PA, TN, TX, VA, WA, WI; and countries of: Belgium, China, and Saudi Arabia.

Product description

Verigene BC-GN Kit (Catalog number 20-005-021); 1) 20 Verigene BC-GN Test Cartridges; are qualitative multiplexed in vitro diagnostic tests for simultaneous detection and identification of selected gram-positive (BC-GP) and gram-negative (BC-GN) bacteria and resistance markers.

Code information (models, lots, serial numbers)

Part Number: 20-009-021; Lot Numbers: 052113021C

Other devices under the same FDA submission

3 devices cleared under K132843 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates