ExacTrac is intended to be used to place patients at an… recallZ-2146-2014 · Brainlab AG · initiated Jun 24, 2014

Class II Terminated All recalls posted in August 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Brainlab AG or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

When using multiple isocenters (radiation treatment targets) within a single plan, in certain workflow conditions ExacTrac v.6.0.4 might move the patient to an unintended isocenter position, despite displaying the green "OK" icon. If this anomaly occurs and is not detected by the user, the radiation treatment dose at the linear accelerator may be delivered to the unintended target position.

FDA root cause category: Software design

What the company is doing

Brainlab sent a FIELD SAFETY NOTICE / PRODUCT NOTIFICATION letter dated June 24, 2014, to all affected customers. The letter included instructions for customers to: 1) Avoid using plans with multiple isocenters in ExacTrac 6.0.4 whenever possible, and 2) If plans with multiple isocenters cannot be avoided and such a plan is used in ExacTrac v.6.0.4, then always select "Yes" in the dialog displayed to "keep the previous correction shift" when switching between isocenters and always verify all final treatment positions. Brainlab will provide a software update with this issue solved to affected customers. Brainlab will actively contact you to schedule the update within the next 6 months. Customers with questions can contact Brainlab at 800-597-5911 or via e-mail at [email protected].

Recall details

Recall number
Z-2146-2014
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 24, 2014
Posted by the FDA
Aug 4, 2014
Recalling firm
Brainlab AG
Quantity
15 systems
Product code
IYE
FDA submissions
K120789
Distribution
Worldwide Distribution - USA including CA, CT, and IL; Internationally to France, Germany, Japan, Netherlands, Russia, and Slovenia.

Product description

ExacTrac is intended to be used to place patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures, in order to treat lesions, tumors and conditions anywhere in the body when radiation treatment is indicated. ExacTrac may also be used to monitor the patient position during the treatment.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

ExacTrac Version 6.0.4; Model/Catalog Numbers: 1) 20833C EXACTRAC 6.0 IR POSITIONING SOFTWARE;  2) 20834C EXACTRAC 6.0 X-RAY POSITIONING SOFTWARE;  3) 20835C EXACTRAC 6.0 IR MONITORING SOFTWARE;  4) 20853C EXACTRAC 6.0 X-RAY LIMITED SOFTWARE;  5) 49926B ET UPGRADE SOFTWARE 3.X TO 6.0 (IR+XR);  6) 49927B ET UPGRADE SOFTWARE 4.X TO 6.0 (IR+XR);  7) 49928B ET UPGRADE SOFTWARE 5.X TO 6.0 (IR+XR);  8) 49933B ET UPGRADE TRUEBEAM 5.5 TO 6.0 LIMITED;  9) 49934A EXACTRAC UPGRADE 5.5 TO 6.0 BASIC;  10) 49936A ET SOFTWARE UPDATE 6.0.X TO 6.0.4;  11) 49938 ET SOFTWARE UPDATE 6.0.3TO 6.0.4;  12) 49939 ET SOFTWARE UPDATE 6.0.X to 6.0.4 P&R;  13) 49973B ET DATA PREP / REVIEW SYSTEM;  14) 49996B ET UPGRADE SOFTWARE X.X TO 6.0 SW ONLY;  15) 49997B ET UPGRADE SOFTWARE 5.5 TO 6.0 LIMITED;  16) 49998B ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT

Other devices under the same FDA submission

2 devices cleared under K120789 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates