ExacTrac is intended to be used to place patients at an… recallZ-2146-2014 · Brainlab AG · initiated Jun 24, 2014
Class II Terminated All recalls posted in August 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Brainlab AG or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
When using multiple isocenters (radiation treatment targets) within a single plan, in certain workflow conditions ExacTrac v.6.0.4 might move the patient to an unintended isocenter position, despite displaying the green "OK" icon. If this anomaly occurs and is not detected by the user, the radiation treatment dose at the linear accelerator may be delivered to the unintended target position.
FDA root cause category: Software design
What the company is doing
Brainlab sent a FIELD SAFETY NOTICE / PRODUCT NOTIFICATION letter dated June 24, 2014, to all affected customers. The letter included instructions for customers to: 1) Avoid using plans with multiple isocenters in ExacTrac 6.0.4 whenever possible, and 2) If plans with multiple isocenters cannot be avoided and such a plan is used in ExacTrac v.6.0.4, then always select "Yes" in the dialog displayed to "keep the previous correction shift" when switching between isocenters and always verify all final treatment positions. Brainlab will provide a software update with this issue solved to affected customers. Brainlab will actively contact you to schedule the update within the next 6 months. Customers with questions can contact Brainlab at 800-597-5911 or via e-mail at [email protected].
Recall details
- Recall number
- Z-2146-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jun 24, 2014
- Posted by the FDA
- Aug 4, 2014
- Recalling firm
- Brainlab AG
- Quantity
- 15 systems
- Product code
- IYE
- FDA submissions
- K120789
- Distribution
- Worldwide Distribution - USA including CA, CT, and IL; Internationally to France, Germany, Japan, Netherlands, Russia, and Slovenia.
Product description
ExacTrac is intended to be used to place patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures, in order to treat lesions, tumors and conditions anywhere in the body when radiation treatment is indicated. ExacTrac may also be used to monitor the patient position during the treatment.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
ExacTrac Version 6.0.4; Model/Catalog Numbers: 1) 20833C EXACTRAC 6.0 IR POSITIONING SOFTWARE; 2) 20834C EXACTRAC 6.0 X-RAY POSITIONING SOFTWARE; 3) 20835C EXACTRAC 6.0 IR MONITORING SOFTWARE; 4) 20853C EXACTRAC 6.0 X-RAY LIMITED SOFTWARE; 5) 49926B ET UPGRADE SOFTWARE 3.X TO 6.0 (IR+XR); 6) 49927B ET UPGRADE SOFTWARE 4.X TO 6.0 (IR+XR); 7) 49928B ET UPGRADE SOFTWARE 5.X TO 6.0 (IR+XR); 8) 49933B ET UPGRADE TRUEBEAM 5.5 TO 6.0 LIMITED; 9) 49934A EXACTRAC UPGRADE 5.5 TO 6.0 BASIC; 10) 49936A ET SOFTWARE UPDATE 6.0.X TO 6.0.4; 11) 49938 ET SOFTWARE UPDATE 6.0.3TO 6.0.4; 12) 49939 ET SOFTWARE UPDATE 6.0.X to 6.0.4 P&R; 13) 49973B ET DATA PREP / REVIEW SYSTEM; 14) 49996B ET UPGRADE SOFTWARE X.X TO 6.0 SW ONLY; 15) 49997B ET UPGRADE SOFTWARE 5.5 TO 6.0 LIMITED; 16) 49998B ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT
Other devices under the same FDA submission
2 devices cleared under K120789 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Exactrac 6.2Brainlab SE · Accelerator, Linear, Medical1 recall6 reports
- Exactrac 6.5Brainlab SE · Accelerator, Linear, Medical1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page