Endo & Unipolar Femoral Heads recallZ-2548-2014 · Zimmer, Inc. · initiated Jul 22, 2014
Class II Terminated All recalls posted in August 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.
FDA root cause category: Package design/selection
What the company is doing
Zimmer sent an URGENT MEDICAL DEVICE RECALL letter dated July 2014 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Please call the Customer Call Center at 1-800-348-2759 for questions or concerns involving this notice.
Recall details
- Recall number
- Z-2548-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 22, 2014
- Posted by the FDA
- Aug 26, 2014
- Recalling firm
- Zimmer, Inc.
- Quantity
- 972,125 units
- Product code
- JDI
- FDA submissions
- K960278
- Distribution
- Worldwide Distribution.
Product description
Endo & Unipolar Femoral Heads: ENDO FEMORAL HEAD 43MM DIA, ENDO FEMORAL HEAD 53MM DIA, ENDO FEMORAL HEAD 47MM DIA, ENDO FEMORAL HEAD 50MM DIA, ENDO FEMORAL HEAD 42MM DIA, ENDO FEMORAL HEAD 54MM DIA, ENDO FEMORAL HEAD 48MM DIA, ENDO FEMORAL HEAD 41MM DIA, ENDO FEMORAL HEAD 42MM DIA, ENDO FEMORAL HEAD 43MM DIA, ENDO FEMORAL HEAD 44MM DIA, ENDO FEMORAL HEAD 45MM DIA, ENDO FEMORAL HEAD 46MM DIA, ENDO FEMORAL HEAD 47MM DIA, ENDO FEMORAL HEAD 48MM DIA, ENDO FEMORAL HEAD 49MM DIA, ENDO FEMORAL HEAD 50MM DIA, ENDO FEMORAL HEAD 51MM DIA, ENDO FEMORAL HEAD 52MM DIA, ENDO FEMORAL HEAD 53MM DIA, ENDO FEMORAL HEAD 54MM DIA, ENDO FEMORAL HEAD 55MM DIA
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Item Numbers : 00781804300, 00781805300, 00781804700, 00781805000, 00781804200, 00781805400, 00781804800, 00781804100, 00781804200, 00781804300, 00781804400, 00781804500, 00781804600, 00781804700, 00781804800, 00781804900, 00781805000, 00781805100, 00781805200, 00781805300, 00781805400, 00781805500 Lot Numbers: 37210958, 37210967, 37211180, 37211183, 37211311, 37211316, 37211320, 37211379, 37211482, 62561484, 37211643, 37211645, 37211646, 37211647, 37211653, 37211655, 37211657, 37211658, 37211661, 37211801, 37211804, 37211805, 37211806, 37211811, 37211812, 37211815, 37211817, 37211820, 37212062, 37212063, 37212065, 37212074, 37212079, 62573574, 62573576, 62576002, 62576008, 62593298, 62593302, 37212608, 37211644
Other devices under the same FDA submission
1 device cleared under K960278 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA 9026-29Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 68818).
- Z-2547-2014 · TerminatedEndo & Unipolar Femoral Heads: ENDO FEMORAL HEAD 41MM DIA ENDO FEMORAL HEAD 42MM DIA ENDO FEMORAL HEAD 43MM DIA ENDO FEMORAL HEAD 44MM DIA ENDO FEMORAL HEAD 45MM DIA ENDO FEMORAL H…
- Z-2549-2014 · TerminatedTrauma Large Screws: CANN BONE SCREW 7.0MMX100MMLG CANN BONE SCREW 7.0X105LG CANN BONE SCREW 7.0X105LG CANN BONE SCREW 7.0MMX105MMLG CANN BONE SCREW 7.0X110LG CANN BONE SCREW 7.0X1…
- Z-2552-2014 · TerminatedKnee Stem Extensions & Impactor Sleeve: STEM IMPLANT 20MMDX145MM STEM IMPLANT 22MMDX145MM STEM IMPLANT 24MMDX145MM SHARP FLUTED ST 20MMDX75MM SHARP FLUTED ST 22MMDX75MM SHARP FLTD …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page