Persona (TM) The Personalized Knee System recallZ-2298-2014 · Zimmer, Inc. · initiated Jun 10, 2013
Class II Terminated All recalls posted in August 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Zimmer had received complaints that reported breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). This is a retrospective recall which covers Zimmer's corrective actions taken earlier this year concerning the TASPs.
FDA root cause category: Device Design
What the company is doing
On February 10, 2014, Zimmer sent letter to surgeons who had filed complaints regarding the affected TASP fractures. The letter provided the surgeons with Zimmer's investigational findings of the issue. Surgeons were directed to send questions or any additional informtion to [email protected]. The Zimmer Weekly Wrap-Up, dated February 14, 2014, was issued electronically to sales force. They were notified to engage in mandatory training to improve the usage of the TASPs and prevent potential breakage. Zimmer has taken additional corrective measures by initiating another recall in July 2014 whereby they notify the distributors as well as all users of the enhanced surgical technique.
Recall details
- Recall number
- Z-2298-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jun 10, 2013
- Posted by the FDA
- Aug 21, 2014
- Recalling firm
- Zimmer, Inc.
- Quantity
- 70,986 distribution events
- Product code
- JWH
- FDA submissions
- K123459
- Distribution
- Worldwide Distribution-USA (nationwide) including DC and Puerto Rico and the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA, WI, WY, and the countries of Australia, Canada, Germany, India, Japan, Korea, and Singapore.
Product description
Persona (TM) The Personalized Knee System CR Tibial Articular Surface Provisional Top Nonsterile TASP CR L 3-9 CD TOP
Code information (models, lots, serial numbers)
Item Numbers: 42517000410 Lot Numbers: 62024975 62043546 62144792 62152725 62165538 62187318 62337961 62372407 62386776 62397159 62405917 62456263 62499170 62551386 62568272 62572572 62565051 62663168 62691384 62707494 62717346
Other devices under the same FDA submission
67 devices cleared under K123459 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Persona® The Personalized Knee® 42-5176-005-05Zimmer, Inc. · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive11 reports
- Persona® The Personalized Knee® 42-5176-007-07Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER11 reports
- PERSONA® Vivacit-E® 42-5226-007-12Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER10 reports
- PERSONA® Vivacit-E® 42-5226-007-10Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER7 reports
- PERSONA® Vivacit-E® 42-5126-005-12Zimmer, Inc. · Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer + additive/metal/polymer + additive6 reports
- PERSONA® Vivacit-E® 42-5126-007-12Zimmer, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER6 reports
- Persona® 42-5279-004-04Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER6 reports
- PERSONA® Vivacit-E® 42-5226-007-14Zimmer, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER5 reports
- PERSONA® Vivacit-E® 42-5126-010-10Zimmer, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER4 reports
- Persona® The Personalized Knee® 42-5176-003-03Zimmer, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER4 reports
- PERSONA® Vivacit-E® 42-5226-005-10Zimmer, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER4 reports
- PERSONA® Vivacit-E® 42-5226-008-10Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER4 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 68817).
- Z-2296-2014 · TerminatedPersona (TM) The Personalized Knee System PS Tibial Articular Surface Provisional Top Nonsterile TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD TOP TASP PS L 3-5 CD T…
- Z-2297-2014 · TerminatedPersona (TM) The Personalized Knee System Tibial Articular Surface Provisional Bottom Nonsterile TASP L CD BOTTOM, +0MM TASP L CD BOTTOM, +6MM TASP L EF BOTTOM, +0MM TASP L EF BO…
- Z-2299-2014 · TerminatedPersona (TM) The Personalized Knee System UC Tibial Articular Surface Provisional Top Nonsterile TASP UC L 3-7 CD TOP TASP UC L 4-11 EF TOP TASP UC L 7-12 GH TOP TASP UC L 9-12 J …
- Z-2300-2014 · TerminatedPersona (TM) The Personalized Knee System Tibial Articular Surface Provisional Bottom Nonsterile PSN A/S PROV L CD BOTTOM +0MM PSN A/S PROV L CD BOTTOM +6MM PSN A/S PROV L EF BOTT…
- Z-2301-2014 · TerminatedPersona (TM) The Personalized Knee System CPS Tibial Articular Surface Provisional Top Nonsterile TASP MLC L 3-5 CD TOP TASP MLC L 6-9 CD TOP TASP MLC L 3-5 EF TOP TASP MLC L 6-9 …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page