The BD MAX MRSA Assay, catalogue #442953 recallZ-2618-2014 · Becton Dickinson & Co. · initiated Jul 23, 2014
Class II Terminated All recalls posted in September 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Becton Dickinson & Co. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The performance of the assay no longer conforms to this statement in the Analytical Specificity section of the Package Insert: One hundred-eleven (111) out of 111 MSSA strains tested at extremely high concentrations (> 106 CFU/swab), produced negative results with the BD MAX MRSA Assay. One of the samples tested with each of the CAP surveys MRS5-A 2014 and MRS5-B 2014 yielded false positive results.
FDA root cause category: Under Investigation by firm
What the company is doing
Becton Dickinson sent an Field Correction Notification initiated on July 23, 2014. US customers were contacted via letters sent by UPS. The letter identified the product the problem and the action needed to be taken by the customer. The acknowledgement form included with the follow up letter that is faxed back to BD will be reconciled with the customer list. BD will contact 10% of non-respondents to verify receipt of the customer letter. Any product returned to BD will be quarantined and discarded. Nonetheless, you may observe some false positive results for MSSA strains when using the above referenced product. Please contact the BD Technical Service Department at 1-800-638-8663 if you have observed an increase in MSSA false positive results or have failed the above referenced CAP surveys.
Recall details
- Recall number
- Z-2618-2014
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jul 23, 2014
- Posted by the FDA
- Sep 10, 2014
- Recalling firm
- Becton Dickinson & Co.
- Quantity
- 585,048 tests
- Product code
- NQX
- FDA submissions
- K120138
- Distribution
- US Distribution including the states of : CA, FL, HI, IL, IN, MD, MI, MN, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA and WI.
Product description
The BD MAX MRSA Assay, catalogue #442953. An automated qualitative in vitro diagnostic test for the direct detection of Methicillin-resistant Staphylococcus aureus (MRSA) DNA from nasal swabs in patie.
Code information (models, lots, serial numbers)
All Lots with the expiration date January 21, 2014 or later
Other devices under the same FDA submission
1 device cleared under K120138 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- BD MAX™ MRSA 442953Geneohm Sciences Canada Inc · SYSTEM, NUCLEIC ACID AMPLIFICATION TEST, DNA, METHICILLIN RESISTANT STAPHYLOCOCCUS AUREUS, DIRECT SPECIMEN1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.