Skip to content

Stago IVD, Liquid Anti-Xa, NDC 3607450003116 recallZ-2650-2014 · Diagnostica Stago, Inc. · initiated Jul 18, 2014

Class III Terminated All recalls posted in September 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Diagnostica Stago, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Diagnostica Stago Inc. received some customer complaints regarding a misreading of Reagent 1 (substrate) barcode on STA analyzers.

FDA root cause category: Under Investigation by firm

What the company is doing

Diagnostica Stago sent a "Quality Information" letter/Customer Acknowledgement form to its customers. The letter identified the product, problem, and actions to be taken by the customers. For additional information and further support contact the Diagnostica Stago Hotline at 800-725-0607.

Recall details

Recall number
Z-2650-2014
Recall class
Class III
Status
Terminated
Initiated by the company
Jul 18, 2014
Posted by the FDA
Sep 16, 2014
Recalling firm
Diagnostica Stago, Inc.
Quantity
514 kits
Product code
KFF
FDA submissions
K111822
Distribution
Distributed USA (nationwide) and the country of Canada.

Product description

Stago IVD, Liquid Anti-Xa, NDC 3607450003116. The STA - Liquid Anti-Xa kits are intended for use with STA-R, STA Compact and STA Satellite Analyzers for the quantitative determination of the plasma levels of unfractionated (UFH) and low molecular weight (LMWH) heparins by measuring their anti-Xa activity on antithrombin in a competitive assay using a synthetic chromogenic substrate.

Code information (models, lots, serial numbers)

510(k) #K111822, Lot #112366, Part # REF 00311US

Other devices under the same FDA submission

2 devices cleared under K111822 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page