External power supply adaptors that accompany Clinitek… recallZ-0081-2015 · Siemens Healthcare Diagnostics, Inc · initiated Sep 19, 2014
Class II Terminated All recalls posted in October 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The external power supply for the Clintek Status analyzer, which is provided separately, is damaged. It can result in an electric shock to the User.
FDA root cause category: Process control
What the company is doing
Siemens sent an Urgent Field Safety Notice dated September 2014, to all by e-mail to all affected Siemens Healthcare Diagnostics Regional offices on September 19, 2014 both in the United States and elsewhere for communication with affected customers. This notice informs customers of the issue with the power supply and the potentially affected products that may contain a damaged power supply. Consignees are asked to complete a Field Correction Effectiveness Check form and return in to Siemens Healthcare Diagnostics via Fax to (312) 275-7795. All affected domestic customers were sent a hard copy of the Urgent Field Safety Notice via Federal Express on 9/23/14. All affected customers outside the US will be provided a copy of an Urgent Field Safety Notice via hard copy, e-mail, and/or fax as determined by each countries local regulations and procedures. Customers with questions were instructed to contact their local Siemens technical support representative. For questions regarding this recall call 877-229-3711.
Recall details
- Recall number
- Z-0081-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Sep 19, 2014
- Posted by the FDA
- Oct 17, 2014
- Recalling firm
- Siemens Healthcare Diagnostics, Inc
- Quantity
- approx. 4200
- Product code
- JIL
- FDA submissions
- K091216
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico and Internationally Canada, Mexico, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Great Britain, Greece, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kazakhstan, Kuwait, Latvia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Qatar, Republic Korea, Romania, Russian Fed., Saudi Arabia, Senegal, Singapore, Slovakia, Slovenia, South Africa, Sweden, Switzerland, Tadjikistan, Taiwan, Turkey, Turkmenistan, Unit.Arab Emir., United Kingdom, Vietnam.
Product description
External power supply adaptors that accompany Clinitek Status+ The Clinitek Status systems is a urine analyzer.
Code information (models, lots, serial numbers)
10379675, 10379676, 10379677, 10379678, 10379679, 10379680, 10379681, 10376324; (Legacy P/N: 1780) Power supply adaptors embossed with the following numerical date codes: 50130, 51130, 52130, 01140, 02140, 03140, 04140, 05140, 06140, 07140, 08140, 09140, 10140, 11140, 12140, 13140, 14140, 15140, 16140, 17140, 18140, 19140, 20140, 21140, 22140, 23140
Other devices under the same FDA submission
2 devices cleared under K091216 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CLINITEK Status®+ Analyzer 10379675Siemens Healthcare Diagnostics Inc. · ENZYMATIC METHOD, CREATININE89 reports
- CLINITEK Status®+ Analyzer 10379677Siemens Healthcare Diagnostics Inc. · DIAZONIUM COLORIMETRY, UROBILINOGEN (URINARY, NON-QUANT.)1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 69228).
- Z-0082-2015 · TerminatedExternal power supply adaptors that accompany Clinitek Status Connect System The Clinitek Status system is a urine analyzer.
- Z-0083-2015 · TerminatedExternal power supply adaptors that accompany Clinitek Status Power Supply Adaptor - Spare Part The Clinitek Status system is a urine analyzer.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page