Skip to content

Siemens Mammomat Inspiration mammography systems recallZ-0086-2015 · Siemens Medical Solutions USA, Inc · initiated Sep 18, 2014

Class II Terminated All recalls posted in October 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The stereo biopsy devices for Mammomat Inspiration mammography systems might have integrated a safety switch which causes a failure of the functionality. The pin implemented in the safety switch may not put enough pressure on the safety circuit to prohibit movement. The needle positioning device may move even with the safety switch being set sideward. If this occurs with the needle being already i

FDA root cause category: Process control

What the company is doing

A safety advisory notice, dated September 18, 2014, was sent to end users that identified the product, problem, and actions to be taken.

Recall details

Recall number
Z-0086-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Sep 18, 2014
Posted by the FDA
Oct 17, 2014
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
3
Product code
MUE
FDA submissions
K122286
Distribution
US Distribution in states of: NY, VA, and MA.

Product description

Siemens Mammomat Inspiration mammography systems. The Mammomat Inspiration system is intended for mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.

Code information (models, lots, serial numbers)

material #10140000, with serial numbers: 5018 3849 5042

Other devices under the same FDA submission

1 device cleared under K122286 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page