HemoCue Glucose 201 Microcuvettes, The HemoCue Glucose 201… recallZ-0873-2015 · Radiometer America Inc · initiated Oct 20, 2014

Class II Terminated All recalls posted in December 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Radiometer America Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Batches of HemoCue Glucose 201 Microcuvettes show discoloration and provide results outside the specification at the end of their shelf life, approx. after 6 months. The HemoCue Glucose 201 Analyzer with plasma conversion multiplies the measured whole blood glucose value by a factor of 1.11 and displays a plasma equivalent glucose result. HemoCue Glucose 201 Microcuvettes are for In Vitro Diag

FDA root cause category: Under Investigation by firm

What the company is doing

The consignees will be notified by a Field Safety Notice. Distributors will be instructed to forward the Field Safety Notice to their end customers and to secure that verification forms will be filled-out by end customers and returned to HemoCue.

Recall details

Recall number
Z-0873-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 20, 2014
Posted by the FDA
Dec 24, 2014
Recalling firm
Radiometer America Inc
Quantity
3,712,900 microcuvettes
Product code
CGA
FDA submissions
K020935
Distribution
Worldwide Distribution-US (nationwide) and the countries of Canada, Austria, AUSTRALIA, SWITZERLAND, COTE D'IVOIRE, CYPRUS, CZECH REPUBLIC, GERMANY, DENMARK, FINLAND, FRANCE, UNITED KINGDOM, GEORGIA, HONG KONG, CROATIA, HUNGARY, IRELAND, INDIA, ICELAND, JAPAN, KENYA, MACEDONIA, MALTA, NETHERLANDS, NORWAY, NEPAL, NEW ZEALAND, POLAND, ROMANIA, RUSSIA, SWEDEN, SLOVENIA, UNITED REPUBLIC OF TANZANIA, and SOUTH AFRICA.

Product description

HemoCue Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the HemoCue Glucose 201 DM Analyzer (hereafter referred to as the HemoCue Glucose 201 Analyzer). HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials. Affected lots identified initially were 110705, 110706, 110715, 110716, 110717, 110718, 110719, and 110720. Subsequent investigation also identified as affected lots 1310227, 1310252, and 1310257. HemoCue Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the HemoCue Glucose 201 DM Analyzer (hereafter referred to as the HemoCue Glucose 201 Analyzer). HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials.

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Model Number 110705:  Lot numbers: 1310500, 1310501, 1310502, 1310503, 1310518, 1310519, 1310520, 1310521, 1310767, 1310768, 1310769, 1310770, 1310771, 1310772, 1310779, 1310780, 1310781, 1310788, 1310789, 1310790, 1310799, 1311524, 1311525, 1311526, 1311527, 1311528, 1311529, 1311530, 1311531, 1311532, 1311533, 1311534, 1311535, 1311536, 1311537, 1311538, 1311539, 1311541, 1311542   Model Number 110706: Lot numbers: 1310207, 1310208, 1310209, 1310210, 1310211, 1310212, 1310213, 1310214, 1310215, 1310216, 1310228, 1310231, 1310232, 1310234, 1310235, 1310238, 1310239, 1310247, 1310249, 1310250, 1310256, 1310257, 1310258, 1310259, 1310260, 1311270, 1311271, 1311272, 1311273, 1311274, 1311279, 1311280, 1311281, 1311282, 1311283, 1311284, 1311291, 1311296, 1311297, 1311314, 1311315, 1311316, 1311317, 1311318,   Model Number 110715: Lot numbers: 1310217, 1310218, 1310219, 1310221, 1310222, 1310229, 1310230, 1310237, 1310240, 1310241, 1310242, 1310251, 1310252, 1311267, 1311268, 1311269, 1311298, 1311301, 1311302, 1311303, 1311304, 1311321, 1311322, 1311323, 1311324  Model Number 110716: Lot numbers: 1310202, 1310203, 1310204, 1310223, 1310224, 1310225, 1310227, 1310243, 1310244, 1310245, 1310246, 1310261, 1311263, 1311264, 1311265, 1311266, 1311277, 1311278, 1311305, 1311306, 1311307, 1311308, 1311309, 1311310, 1311311, 1311312, 1311313, 1311325   Model Number 110717: Lot numbers: 1310504, 1310505, 1310506, 1310507, 1310508, 1310509, 1310510, 1310514, 1310515, 1310516, 1310517, 1310773, 1310774, 1310775, 1310776, 1310783, 1310784, 1310785, 1310791, 1310792, 1310794, 1310795, 1310796, 1311540, 1311543, 1311544, 1311545   Model Number 110718: Lot numbers: 1310511, 1310512, 1310513, 1310777, 1310778, 1310786, 1310787, 1310797, 1310798  Model Number 110719: Lot numbers: 1310236, 1310237, 1310248, 1311285, 1311286, 1311287, 1311288, 1311289, 1311290, 1311292, 1311293, 1311294, 1311295, 1311299, 1311300  Model Number 110720: Lot numbers: 1310205, 1310206, 1310226, 1310253, 1310254, 1310255, 1311319, 1311320

Other devices under the same FDA submission

1 device cleared under K020935 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.