Skip to content

Mammomat Inspiration system recallZ-0523-2015 · Siemens Medical Solutions USA, Inc · initiated Oct 28, 2014

Class II Terminated All recalls posted in December 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

It was determined that if the Mammomat Inspiration system is not secured to the floor (per customer request) and monitors are positioned too far from the table top, there is a potential risk that the Acquisition Workstation (AWS) table may become unstable and fall over. This may result in a serious injury to the operator.

FDA root cause category: Use error

What the company is doing

A safety advisory notice, dated October 28, 2014, was sent to users that identified the product, problem, and actions to be taken to prevent risk to patients and users. Siemens technicians werre instructed to contact customers to schedule a field modification.

Recall details

Recall number
Z-0523-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 28, 2014
Posted by the FDA
Dec 6, 2014
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
3
Product code
MUE
FDA submissions
K122286
Distribution
US Nationwide Distribution in the state of: WV, NY, and FL..

Product description

Mammomat Inspiration system: Product Usage: mammography exams, screening, diagnosis, and stereotactic biopsies under the supervision of medical professionals. Mammographic images can be interpreted by either hard copy film or soft copy workstation.

Code information (models, lots, serial numbers)

model number: 10140000

Other devices under the same FDA submission

1 device cleared under K122286 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page