LEGION(TM) COCR CONSTRAINED FEMORAL COMPONENT, SIZE 6… recallZ-0591-2015 · Smith & Nephew, Inc. · initiated Nov 4, 2014
Class II Terminated All recalls posted in December 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smith & Nephew, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The stem taper on this device is undersized; the gauge point is not deep enough.
FDA root cause category: Process control
What the company is doing
The firm, Smith & Nephew, Inc., sent an "Urgent-Product Recall 1st Notification-Urgent" letter dated 11/4/ 2014 to its customers via E-mail on 11/4/2014. International distributors were notified on 11/4/2014 by Email. Non-responding consignees will be contacted on 11/25/2014 by original method. The letter described the product, problem, and actions to be taken. The customers were instructed to inspect your inventory and locate any unused devices and immediately quarantine; complete and return the Inventory Return Certification Form via fax to: +1-901-566-7975, even if you do not have product to return; contact Smith & Nephew's Safety Affairs Department via email at [email protected] or fax +1-901-566-7975 to obtain return authorization (RA) number; and return affected product to Smith & Nephew, Attn: Global Field Actions, 3303 E Holmes Road, Memphis, TN 38118. If you have any questions, please contact the Group Director, Quality Assurance +1-901-399-1970.
Recall details
- Recall number
- Z-0591-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 4, 2014
- Posted by the FDA
- Dec 12, 2014
- Recalling firm
- Smith & Nephew, Inc.
- Quantity
- 23 units
- Product code
- JWH
- FDA submissions
- K060742
- Distribution
- Worldwide Distribution: US (nationwide) including states of: MI, MN, MO, OH, PA, IA, and TN; and Internationally to: SWITZERLAND; CANADA, COLOMBIA, GERMANY, Dubai, (UAE), THAILAND; SOUTH AFRICA, SOUTH KOREA; and SINGAPORE.
Product description
LEGION(TM) COCR CONSTRAINED FEMORAL COMPONENT, SIZE 6, LEFT, REF 71426006, QTY 1, Smith & Nephew, knee prosthesis Usage:orthopedic.
Code information (models, lots, serial numbers)
Lot Number 14EM23391
Other devices under the same FDA submission
36 devices cleared under K060742 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- LEGION 71425005Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER16 reports
- LEGION 71425007Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER14 reports
- LEGION 71426006Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER12 reports
- LEGION 71425006Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER11 reports
- LEGION 71426005Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER10 reports
- LEGION 71423235Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER9 reports
- LEGION 71426004Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER9 reports
- LEGION 71423236Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER7 reports
- LEGION 71425008Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER7 reports
- LEGION 71426007Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER7 reports
- LEGION 71423237Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER6 reports
- LEGION 71423225Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER6 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 69759).
- Z-0589-2015 · TerminatedLEGION(TM) COCR CONSTRAINED FEMORAL COMPONENT, SIZE 4, LEFT, REF 71425004, QTY 1, Smith & Nephew, knee prosthesis Usage:orthopedic.
- Z-0590-2015 · TerminatedLEGION(TM) COCR CONSTRAINED FEMORAL COMPONENT, SIZE 5, RIGHT, REF 71426005, QTY 1, Smith & Nephew, knee prosthesis Usage:orthopedic.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page