Skip to content

MEVION S250, used for proton radiation therapy recallZ-0588-2015 · Mevion Medical Systems, Inc. · initiated Nov 10, 2014

Class II Terminated All recalls posted in December 2014

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Mevion Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software defect that causes an incorrect dose compensation function to be applied to the internal dose ionization chamber when either pressure or temperature sensor malfunction. This could result in an incorrect dose delivery of no more than 5%.

FDA root cause category: Software design

What the company is doing

Mevion Medical Systems sent an Important Safety Notice letter to the affected customer. The letter identified the affected product, problem and actions to be taken. A software update addressing the issue will be provided to the customer on or about December 5, 2014. For questions or concerns, email [email protected] or call 978-540-1555.

Recall details

Recall number
Z-0588-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Nov 10, 2014
Posted by the FDA
Dec 12, 2014
Recalling firm
Mevion Medical Systems, Inc.
Quantity
1
Product code
LHN
FDA submissions
K120676
Distribution
US Distribution in the state of MO

Product description

MEVION S250, used for proton radiation therapy.

Code information (models, lots, serial numbers)

Serial Number: S250-0001

Other devices under the same FDA submission

1 device cleared under K120676 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page