Trilogy Fiber Metal Acetabular Shell Size 54 mm OD recallZ-0875-2015 · Zimmer Manufacturing B.V. · initiated Dec 4, 2014
Class II Terminated All recalls posted in December 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Manufacturing B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Zimmer received a single complaint wherein a Trilogy® Fiber Metal Multi-Hole Acetabular Shell would not attach to the inserter/impactor instrumentation. Subsequent investigation found the polar-hole threads to be present but non-conforming to specifications, in that the minor diameter was undersized. Investigation also indicates that the issue is isolated and affects only lot 62714305, which
FDA root cause category: Under Investigation by firm
What the company is doing
The firm, Zimmer, sent an "URGENT MEDICAL DEVICE RECALL-LOT SPECIFIC" memo dated 12/4/2014 to its customers. The memo described the product, problem and actions to be taken. The customers were instructed to locate all product from the listed lot and quarantine them immediately; carry out a physical count of all affected product in inventory; complete and return the Inventory Return Certification Form via email to: [email protected] on or before 12/12/2014; return the recalled product along with Inventory Return Certification Form on or before 12/12/2014; keep and maintain a copy for your records; and notify Zimmer of any hospitals that you have further distributed the affected product to and complete spreadsheet template provided. After reviewing this notification you have further questions or concerns, call the customer call center at 1-800-348-2759 between 8:00 am and 5:00 pm EST.
Recall details
- Recall number
- Z-0875-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 4, 2014
- Posted by the FDA
- Dec 24, 2014
- Recalling firm
- Zimmer Manufacturing B.V.
- Quantity
- 68 units
- Product code
- LPH
- FDA submissions
- K934765
- Distribution
- Worldwide Distribution: US (nationwide) and countries of: France, Germany, Iran, Romania, Serbia, Singapore, Taiwan, and United Kingdom.
Product description
Trilogy Fiber Metal Acetabular Shell Size 54 mm OD The Trilogy Acetabular System is indicated for either cemented or uncemented use in skeletally mature individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.
Code information (models, lots, serial numbers)
Item No. 00-6200-054-20; Lot 62714305
Other devices under the same FDA submission
8 devices cleared under K934765 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Trilogy® 6200-56-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED5 reports
- Trilogy® 6200-50-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 reports
- Trilogy® 6200-52-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- Trilogy® 6200-60-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- Trilogy® 6200-54-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Trilogy® 6200-62-20Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Trilogy® 6200-66-22Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Trilogy® 6105-50-32Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page