ADULT Radiotransparent Electrode, Part number C100-PHILIPS… recallZ-0583-2015 · Heart Sync, Inc. · initiated Nov 11, 2014
Class I Terminated All recalls posted in December 2014
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Heart Sync, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Heart Sync Multi-function defibrillation electrodes will not connect with Philips FR3 or FRx AED units. The FRx AED unit requires the pads to be pre-connected, and will issue a continuous alarm chirp to alert the user that the proper pads are not connected to the unit prior to use. The FR3, however, does not require pre-connection and the user may not discover the incompatibility issue until the
FDA root cause category: Device Design
What the company is doing
The firm, HeartSync, sent an "Voluntary Field Safety Alert for Heart Sync Multi-function Defibrillator Electrodes" notice dated 11/11/2014 to customers via email notifying them of the incompatibility with the Philips FR3 and FRX AEDs and requesting response to the notice. On 12/1/2014 (dated 11/26/2014) Heart Sync issued a Voluntary Device Correction via Press Release. The Press Release included a description of the reason for the recall, affected product, and instruction that the Philips FR3 and FRx AED units should only be used with electrodes specified in the (Philips) equipment manuals. For further information or to report a problem, please contact Heart Sync at 734-213-5530, 24 hours a day, 7 days a week, ,or email at [email protected].
Recall details
- Recall number
- Z-0583-2015
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Nov 11, 2014
- Posted by the FDA
- Dec 24, 2014
- Recalling firm
- Heart Sync, Inc.
- Quantity
- sold in boxes of ten electrodes: 4663.8 boxes; 46638 units total
- Product code
- MKJ
- FDA submissions
- K080421
- Distribution
- Worldwide Distribution - USA (Nationwide) in the states of NM, VT, MN, IL, MO, OH, MI, OR, ND, KY, PA, TX, WA, FL, NE, AR, VA, DE, GA, NY, CA, TN, AZ, RI, MA, and PR; and countries of: Thailand and South Korea.
Product description
ADULT Radiotransparent Electrode, Part number C100-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector", Intended for use during defibrillation, cardioversion, pacing, and ECG monitoring, Single use.
Code information (models, lots, serial numbers)
All Lots. Catalog / Part Number C100-PHILIPS
Other devices under the same FDA submission
4 devices cleared under K080421 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Curaplex 16384Bound Tree Medical, LLC · Electrode, electrodcardiograph, multi-function4 reports
- Curaplex 16242Bound Tree Medical, LLC · Electrode, electrodcardiograph, multi-function1 report
- Curaplex 16250Bound Tree Medical, LLC · Electrode, electrodcardiograph, multi-function1 report
- Curaplex 16381Bound Tree Medical, LLC · Electrode, electrodcardiograph, multi-function1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 69875).
- Z-0584-2015 · TerminatedADULT/CHILD Radiotransparent Electrode, Part number C100AC-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Co…
- Z-0585-2015 · TerminatedADULT Radiotranslucent Electrode, Part number T100LO-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connecto…
- Z-0586-2015 · TerminatedADULT Radiotranslucent Electrode, Part number T100-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Connector",…
- Z-0587-2015 · TerminatedADULT/CHILD Radiotranslucent Electrode, Part number T100AC-PHILIPS, Rx ONLY, For use with Philips HeartStart AED "Comparable to Philips Medical HeartStart Electrodes Plug Style Con…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page