BIOLOX DELTA recallZ-0825-2015 · DePuy Orthopaedics, Inc. · initiated Nov 7, 2014

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from DePuy Orthopaedics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Two complaints reported that the BIOLOX delta TS Ceramic Femoral Articul/eze12/14 36mm +12 Head (product code 1365-36-740) did not fit like the trial; the definite hip construct felt loose. The loosening refers to tissue tension and not the taper connection between the femoral head and stem. The investigations found that the small size sleeve (+1.5 femoral head offset) was incorrectly used in the

FDA root cause category: Mixed-up of materials/components

What the company is doing

On 11/7/2014, URGENT INFORMATION - RECALL NOTICE notifications were sent to the affected distributors, hospitals, and doctors with instructions for returning the affected product. All distributors were notified via electronic mail, and all hospitals and doctors were notified via sales consultant visit. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. For clinical questions from surgeons, please contact DePuy Orthopaedics, Inc. Scientific Information Office at 1-888-554-2482 (M-F; 8 a.m. - 5 p.m. EDT). For questions about device recall information provided, please contact Kim Earle, Recall Coordinator, at 574-371-4917 (M-F; 8 a.m. - 5 p.m. EDT).

Recall details

Recall number
Z-0825-2015
Status
Terminated
Initiated by the company
Nov 7, 2014
Posted by the FDA
Dec 17, 2014
Recalling firm
DePuy Orthopaedics, Inc.
Quantity
19 units worldwide
Product code
LZO
FDA submissions
K071830
Distribution
Distributed in the states of WV, AZ, and AL, and the countries of Germany, and Finland.

Product description

Ceramic Femoral Head. Intended for use in total hip arthroplasty applications to replace the articular surface of the femoral head in primary hip surgery and for the salvage of a failed previous hip surgery.

Code information (models, lots, serial numbers)

Lot 7928614 CATALOG NO. : 1365-36-740 Barcode GTIN: 10603295033684