Trident PSL HA Solid Back 50 mm recallZ-1043-2015 · Stryker Howmedica Osteonics Corp. · initiated Dec 15, 2014
Class II Terminated All recalls posted in February 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Howmedica Osteonics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Five customer complaints were received for the same batch of Trident PSL Shells (part number 540-11-50E lot 49073501) alleging that when the packages were opened there were no dome holes plugs present.
FDA root cause category: Process control
What the company is doing
Stryker sent an Urgent Medical Device letter dated January 7, 2015, with a Recall Notification/Acknowledgement Form via Fed Ex to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to complete and return the attached Product Recall Acknowledgement Form to [email protected] or fax to 1-855-251-3635. Customers should return the affected product to: Attention: Regulatory Compliance Stryker Orthopaedics 325 Corporate Drive Mahwah, New Jersey 07430 Customers with questions should call 201-831-6365. For questions regarding this recall call 201-831-5000.
Recall details
- Recall number
- Z-1043-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Dec 15, 2014
- Posted by the FDA
- Feb 3, 2015
- Recalling firm
- Stryker Howmedica Osteonics Corp.
- Quantity
- seven units
- Product code
- MEH
- FDA submissions
- K983382
- Distribution
- Nationwide Distribution including NY, OH, PA, and TX.
Product description
Trident PSL HA Solid Back 50 mm Includes Dome Hole Plug Ref 540-11-50E Intended for cementless application in a total hip arthroplasty procedure. This system is a two piece ( a shell and a liner) design that is assembled during surgery. The shell is press-fit into the patients prepared acetabulum and a linger (polyethylene or alumina ceramic) is then locked into the shell. The femoral head component of the total hip system articulates with the acetabular liner.
Code information (models, lots, serial numbers)
510K 983382 Lot code 49073501
Other devices under the same FDA submission
9 devices cleared under K983382 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Trident 621-00-28EHowmedica Osteonics Corp. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 reports
- Trident 621-00-28GHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate1 report
- Trident 621-00-28IHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate1 report
- Trident 621-10-26EHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate1 report
- Trident 621-10-26DHowmedica Osteonics Corp. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Trident 621-10-26FHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate1 report
- Trident 621-10-28EHowmedica Osteonics Corp. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Trident 621-10-28FHowmedica Osteonics Corp. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
- Trident 621-10-28GHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page