BD Viper LT System, catalog number 442839 when used in… recallZ-1816-2015 · Becton Dickinson & Co. · initiated Nov 7, 2014
Class II Terminated All recalls posted in June 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Becton Dickinson & Co. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for false positive Chlamydia trachomatis (CT) results and Neisseria gonorrhoeae (GC) results when processing the CT Qx and GC Qx Assays on the BD Viper LT Instrument.
FDA root cause category: Device Design
What the company is doing
BD sent an Urgent Field Corrective Action letter dated November 2014, to all affected customers. The letter identified the problem the product and the action needed to be taken by the customer. Customers were advised: a. Include the date recall was initiated, if it is already underway or the date your firm plans to start the recalling process. b. How do you plan to notify all the consignees affected by this recall? (press release, letter, telefax, telephone, e-mail, visit, etc.) The customer was initially contacted via a phone call and subsequently provided with a written communication. c. How do you plan to monitor the number of consignees non-responding to the recall communication? The single US customer was contacted by phone and provided the written communication. d. How do you plan to do effectiveness checks of this recall? The affected instrument will be corrected. e. Date your firm ceased further distribution of the product(s). Product was not placed on hold. This issue does not affect new customers. f. How do you plan to store the recall product? BD has not requested that instruments or reagents be returned. g. How do you plan to dispose of the recall products? BD is not requesting that customers return the instruments or reagents. If you require further assistance or clarification, please contact the BD Technical Services Department at 1-800 638-8663.
Recall details
- Recall number
- Z-1816-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Nov 7, 2014
- Posted by the FDA
- Jun 18, 2015
- Recalling firm
- Becton Dickinson & Co.
- Quantity
- 7 Viper LT Systems
- Product code
- OOI
- FDA submissions
- K140447
- Distribution
- Worldwide distribution to US (state of LA), Hong Kong, Austria, China, France, and the UK.
Product description
BD Viper LT System, catalog number 442839 when used in conjunction with the BD ProbeTec Chlamydia trachomatis (CT) Qx Assay (cat. #442959) and the BD ProbeTec Neisseria gonorrhoeae (GC) Qx Assay (cat. #442842). In vitro diagnostic.
Code information (models, lots, serial numbers)
Only the BD Viper LT Systems, catalog number 442839, that utilize the CT/GC Assays are affected. Serial numbers VLT0040, VLT0001, VLT0015, VLT0016, VLT0023, VLT0026, VLT0033.
Other devices under the same FDA submission
1 device cleared under K140447 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- BD Viper™ LT System 442839Becton, Dickinson And Company · real time Nucleic acid amplification system37 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.