Skip to content

Persona (TASP) Tibial Articular Surface Provisional Shim… recallZ-1052-2015 · Zimmer, Inc. · initiated Dec 11, 2014

Class II Terminated All recalls posted in February 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for delay in surgery and/or ball bearings to be left in the wound. Repeated cycles of sonic cleaning can break down the ball bearing retention feature (swage) on the shim. To prevent this, TASP shims must be manually cleaned and inspected per product insert, 87-6203-991-22.

FDA root cause category: Device Design

What the company is doing

Zimmer sent an URGENT MEDICAL DEVICE CORRECTION notifications letter dated December 11, 2014 to affected distributors via electronic mail, and all hospital Risk Managers and distributors with affected inventory via courier. The letter identified the affected product, problem, instructions for responding to the formal recall notification.and actions to be taken. For questions call the customer call center at 1-800-348-2759.

Recall details

Recall number
Z-1052-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Dec 11, 2014
Posted by the FDA
Feb 4, 2015
Recalling firm
Zimmer, Inc.
Quantity
67963
Product code
OIY
FDA submissions
K123459
Distribution
Worldwide Distribution - USA Nationwide including DC, and PR and the countries of Australia, Austria, Belgium, Canada, Czech Republic, Dominican Republic, France, Germany, India, Iran, Israel, Italy, Japan, Luxembourg, Netherlands, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.

Product description

Persona (TASP) Tibial Articular Surface Provisional Shim, all sizes and Lots Catalog Number: 42-5279-002-00 thru 42-5279-009-04. Product Usage: The Persona TASPs are sterilizable instruments Intended for multiple uses. The TASP components are used to aid the surgeon in assessing the thickness of the articular surface required. The TASP construct is utilized while conducting a trial range of motion. The TASP consists of 3 parts: a TASP bottom, a TASP shim, and a TASP top. The shims come in varying thicknesses to allow trialing for specific Articular Surface implant thicknesses.

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Persona (TASP) Tibial Articular Surface Provisional Shim, all sizes and Lots Catalog Number: 42-5279-002-00 thru 42-5279-009-04: 42-5279-002-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, CD, 10mm; 42-5279-002-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, CD, 12mm; 42-5279-002-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, CD, 14mm; 42-5279-004-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, EF, 10mm; 42-5279-004-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, EF, 12mm; 42-5279-004-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, EF, 14mm; 42-5279-006-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, GH, 10mm; 42-5279-006-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, GH, 12mm; 42-5279-006-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, GH, 14mm; 42-5279-008-00 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, J, 10mm; 42-5279-008-02 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, J, 12mm; 42-5279-008-04 All Lots Persona Constrained TASP (Tibial A/S Prov.) Shim, J, 14mm; 42-5279-003-00 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 10mm; 42-5279-003-01 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 11mm; 42-5279-003-02 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 12mm; 42-5279-003-03 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 13mm; 42-5279-003-04 All Lots Persona TASP (Tibial A/S Provisional) Shim, CD, 14mm; 42-5279-005-00 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 10mm; 42-5279-005-01 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 11mm; 42-5279-005-02 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 12mm; 42-5279-005-03 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 13mm; 42-5279-005-04 All Lots Persona TASP (Tibial A/S Provisional) Shim, EF, 14mm; 42-5279-007-00 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 10mm; 42-5279-007-01 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 11mm; 42-5279-007-02 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 12mm; 42-5279-007-03 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 13mm; 42-5279-007-04 All Lots Persona TASP (Tibial A/S Provisional) Shim, GH, 14mm.

Other devices under the same FDA submission

67 devices cleared under K123459 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page